Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications

Francesca Saveria Maddaloni iD, Antonio Sarno iD, Alessandro Loria iD, Anna Piai iD, Cristina Lenardi, Antonio Esposito, Antonella Del Vecchio iD

Sensors · 2025

DOI: 10.3390/s25237338 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Francesca Saveria Maddaloni und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics 'Aldo Pontremoli', University of Milan, 20133 Milan, Italy. Hauptthemen laut MEDLINE: Whole Body Imaging, Phantoms, Imaging, Tomography, X-Ray. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Photon-counting computed tomography (PCCT) represents a major technological advance in clinical CT imaging, offering superior spatial resolution, enhanced material discrimination, and potential radiation dose reduction compared to conventional energy-integrating detector systems. As the first clinically approved PCCT scanner becomes available, establishing a comprehensive characterization of its image quality is essential to understand its performance and clinical impact.

Methods

Image quality was evaluated using a commercial quality assurance phantom with acquisition protocols typically used for three anatomical regions—head, abdomen/thorax, and inner ear—representing diverse clinical scenarios. Each region was scanned using both ultra-high-resolution (UHR, 120 × 0.2 mm slices) and conventional (144 × 0.4 mm slices) protocols. Conventional metrics, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), slice thickness accuracy, and uniformity, were assessed following international standards. Task-based analysis was also performed through target transfer function (TTF), noise power spectrum (NPS), and detectability index (d′) to evaluate diagnostic relevance.

Results

UHR protocols provided markedly improved spatial resolution, particularly in the inner ear imaging, as confirmed by TTF analysis, though with increased noise and reduced low-contrast detectability in certain conditions. CT numbers showed linear correspondence with known attenuation coefficients across all protocols.

Conclusions

This study establishes a detailed technical characterization of the first clinical PCCT scanner, demonstrating significant improvements in terms of spatial resolution and accuracy of the quantitative image analysis, while highlighting the need for noise–contrast optimization in high-resolution imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Whole Body ImagingPhantoms, ImagingTomography, X-RayHumansSignal-To-Noise RatioImage Processing, Computer-AssistedLinear ModelsPhotons

Stichwörter der Autor:innen

CT image qualityphoton-counting detectorstask-based metrics

Autor:innen und Institutionen

  • Department of Physics 'Aldo Pontremoli', University of Milan, 20133 Milan, Italy.
  • Milano Division, National Institute for Nuclear Physics, 20133 Milan, Italy.
  • Medical Phyisics Department, IRCCS San Raffaele Hospital, 20132 Milan, Italy.
  • Experimental Imaging Center, Vita-Salute San Raffaele University, 20132 Milan, Italy.

ORCID-Profile: Francesca Saveria Maddaloni, Antonio Sarno, Alessandro Loria, Anna Piai, Antonella Del Vecchio

Zitieren

APA

Maddaloni, F. S., Sarno, A., Loria, A., Piai, A., Lenardi, C., Esposito, A., & Del Vecchio, A. (2025). Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications. Sensors. https://doi.org/10.3390/s25237338

Vancouver

Maddaloni FS, Sarno A, Loria A, Piai A, Lenardi C, Esposito A, et al. Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications. Sensors. 2025. doi:10.3390/s25237338

BibTeX
@article{Maddaloni2025,
  author = {Maddaloni, Francesca Saveria and Sarno, Antonio and Loria, Alessandro and Piai, Anna and Lenardi, Cristina and Esposito, Antonio and Del Vecchio, Antonella},
  title = {Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25237338}
}
Kostenlos starten →