EUVIMEDAtklāt EUVIMED
Open access · CC BY 4.0

Force Sensor for Instrumented Patellar Prostheses: Development and Characterization

Vera Maioli iD, Matteo Zauli iD, Angelo Cappello iD, Luca Cristofolini iD

Sensors · 2025

DOI: 10.3390/s25041226 ↗

At a glance

Journal article in Sensors, 2025. Written by Vera Maioli and 3 co-authors. Institution: Department of Industrial Engineering, School of Engineering and Architecture, Alma Mater Studiorum, University of Bologna, Via Umberto Terracini 24-2…. Major MEDLINE topics: Patella, Knee Prosthesis, Biosensing Techniques. Cited 4 times so far (Crossref); 21 references. The full text is freely available (open access, CC BY 4.0).

AbstraktsSubjectsBibliogrāfiskie datiCiteRelated studies

Abstract (original language)

The development of an instrumented patellar prosthesis, able to measure the contact forces at the patellofemoral joint, can significantly aid in investigating the causes of total knee arthroplasty failures due to patellar complications. This study focuses on developing and validating an instrumented patellar prosthesis to measure contact forces in the patellofemoral joint. A piezoresistive force sensor was characterized and integrated into a conditioning circuit, with the aim of its implementation in the prosthesis. To measure medial and lateral forces independently, the sensors were trimmed in half. Compression tests (up to 2000 N) assessed sensor performance in terms of linearity (R2 = 0.998 intact vs. 0.989 trimmed), repeatability (0.9% intact vs. 0.8% trimmed), and accuracy (1.7% intact vs. 2.3% trimmed) for forces up to 250 N. Higher force levels resulted in increased errors, but at a rate still comparable to that of existing sensors in the literature. Key considerations for the design of the instrumented prosthesis, such as minimizing point and shear loads, were identified. A prototype prosthesis capable of housing the sensor was proposed. The integrated system shows potential for improving the understanding of Total knee arthroplasty (TKA) failures through in vitro studies and could serve as an intraoperative tool for the evaluation of bone resections.

Subjects

MeSH terms (MEDLINE)

PatellaKnee ProsthesisBiosensing TechniquesHumansArthroplasty, Replacement, KneeProsthesis DesignBiomechanical PhenomenaPatellofemoral JointStress, Mechanical

Author keywords

force sensorinstrumented prosthesispatellapatellofemoral jointpiezoresistivetotal knee arthroplasty

Authors and institutions

  • Department of Industrial Engineering, School of Engineering and Architecture, Alma Mater Studiorum, University of Bologna, Via Umberto Terracini 24-28, 40131 Bologna, Italy.
  • CIRI ICT, Alma Mater Studiorum, University of Bologna, Viale Risorgimento 2, 40136 Bologna, Italy.
  • Department of Electrical, Electronic and Information Engineering, Alma Mater Studiorum, University of Bologna, Viale Risorgimento 2, 40136 Bologna, Italy.

ORCID profiles: Vera Maioli, Matteo Zauli, Angelo Cappello, Luca Cristofolini

Cite

APA

Maioli, V., Zauli, M., Cappello, A., & Cristofolini, L. (2025). Force Sensor for Instrumented Patellar Prostheses: Development and Characterization. Sensors. https://doi.org/10.3390/s25041226

Vancouver

Maioli V, Zauli M, Cappello A, Cristofolini L. Force Sensor for Instrumented Patellar Prostheses: Development and Characterization. Sensors. 2025. doi:10.3390/s25041226

BibTeX
@article{Maioli2025,
  author = {Maioli, Vera and Zauli, Matteo and Cappello, Angelo and Cristofolini, Luca},
  title = {Force Sensor for Instrumented Patellar Prostheses: Development and Characterization},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25041226}
}
Start for free →