Force Sensor for Instrumented Patellar Prostheses: Development and Characterization

Vera Maioli iD, Matteo Zauli iD, Angelo Cappello iD, Luca Cristofolini iD

Sensors · 2025

DOI: 10.3390/s25041226 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Vera Maioli und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Industrial Engineering, School of Engineering and Architecture, Alma Mater Studiorum, University of Bologna, Via Umberto Terracini 24-2…. Hauptthemen laut MEDLINE: Patella, Knee Prosthesis, Biosensing Techniques. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The development of an instrumented patellar prosthesis, able to measure the contact forces at the patellofemoral joint, can significantly aid in investigating the causes of total knee arthroplasty failures due to patellar complications. This study focuses on developing and validating an instrumented patellar prosthesis to measure contact forces in the patellofemoral joint. A piezoresistive force sensor was characterized and integrated into a conditioning circuit, with the aim of its implementation in the prosthesis. To measure medial and lateral forces independently, the sensors were trimmed in half. Compression tests (up to 2000 N) assessed sensor performance in terms of linearity (R2 = 0.998 intact vs. 0.989 trimmed), repeatability (0.9% intact vs. 0.8% trimmed), and accuracy (1.7% intact vs. 2.3% trimmed) for forces up to 250 N. Higher force levels resulted in increased errors, but at a rate still comparable to that of existing sensors in the literature. Key considerations for the design of the instrumented prosthesis, such as minimizing point and shear loads, were identified. A prototype prosthesis capable of housing the sensor was proposed. The integrated system shows potential for improving the understanding of Total knee arthroplasty (TKA) failures through in vitro studies and could serve as an intraoperative tool for the evaluation of bone resections.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PatellaKnee ProsthesisBiosensing TechniquesHumansArthroplasty, Replacement, KneeProsthesis DesignBiomechanical PhenomenaPatellofemoral JointStress, Mechanical

Stichwörter der Autor:innen

force sensorinstrumented prosthesispatellapatellofemoral jointpiezoresistivetotal knee arthroplasty

Autor:innen und Institutionen

  • Department of Industrial Engineering, School of Engineering and Architecture, Alma Mater Studiorum, University of Bologna, Via Umberto Terracini 24-28, 40131 Bologna, Italy.
  • CIRI ICT, Alma Mater Studiorum, University of Bologna, Viale Risorgimento 2, 40136 Bologna, Italy.
  • Department of Electrical, Electronic and Information Engineering, Alma Mater Studiorum, University of Bologna, Viale Risorgimento 2, 40136 Bologna, Italy.

ORCID-Profile: Vera Maioli, Matteo Zauli, Angelo Cappello, Luca Cristofolini

Zitieren

APA

Maioli, V., Zauli, M., Cappello, A., & Cristofolini, L. (2025). Force Sensor for Instrumented Patellar Prostheses: Development and Characterization. Sensors. https://doi.org/10.3390/s25041226

Vancouver

Maioli V, Zauli M, Cappello A, Cristofolini L. Force Sensor for Instrumented Patellar Prostheses: Development and Characterization. Sensors. 2025. doi:10.3390/s25041226

BibTeX
@article{Maioli2025,
  author = {Maioli, Vera and Zauli, Matteo and Cappello, Angelo and Cristofolini, Luca},
  title = {Force Sensor for Instrumented Patellar Prostheses: Development and Characterization},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25041226}
}