EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Simultaneous Dry and Gel-Based High-Density Electroencephalography Recordings

Patrique Fiedler iD, Uwe Graichen iD, Ellen Zimmer, Jens Haueisen iD

Sensors · 2023

DOI: 10.3390/s23249745 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Patrique Fiedler und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Biomedical Engineering and Informatics, Technische Universität Ilmenau, 98693 Ilmenau, Germany. Hauptthemen laut MEDLINE: Evoked Potentials, Visual, Electroencephalography. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Federal Ministry of Education and Research, Freistaat Thüringen, European Union.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Evaluations of new dry, high-density EEG caps have only been performed so far with serial measurements and not with simultaneous (parallel) measurements. For a first comparison of gel-based and dry electrode performance in simultaneous high-density EEG measurements, we developed a new EEG cap comprising 64 gel-based and 64 dry electrodes and performed simultaneous measurements on ten volunteers. We analyzed electrode–skin impedances, resting state EEG, triggered eye blinks, and visual evoked potentials (VEPs). To overcome the issue of different electrode positions in the comparison of simultaneous measurements, we performed spatial frequency analysis of the simultaneously measured EEGs using spatial harmonic analysis (SPHARA). The impedances were 516 ± 429 kOhm (mean ± std) for the dry electrodes and 14 ± 8 kOhm for the gel-based electrodes. For the dry EEG electrodes, we obtained a channel reliability of 77%. We observed no differences between dry and gel-based recordings for the alpha peak frequency and the alpha power amplitude, as well as for the VEP peak amplitudes and latencies. For the VEP, the RMSD and the correlation coefficient between the gel-based and dry recordings were 1.7 ± 0.7 μV and 0.97 ± 0.03, respectively. We observed no differences in the cumulative power distributions of the spatial frequency components for the N75 and P100 VEP peaks. The differences for the N145 VEP peak were attributed to the different noise characteristics of gel-based and dry recordings. In conclusion, we provide evidence for the equivalence of simultaneous dry and gel-based high-density EEG measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Evoked Potentials, VisualElectroencephalographyHumansReproducibility of ResultsElectrodesElectric Impedance

Stichwörter der Autor:innen

EEGbrain imagingdry electrodeselectroencephalographyspatial frequenciesspatial harmonicsvalidation

Autor:innen und Institutionen

  • Institute of Biomedical Engineering and Informatics, Technische Universität Ilmenau, 98693 Ilmenau, Germany.
  • Department of Biostatistics and Data Science, Karl Landsteiner University of Health Sciences, 3500 Krems an der Donau, Austria.

ORCID-Profile: Patrique Fiedler, Uwe Graichen, Jens Haueisen

Förderung

  • Federal Ministry of Education and Research
  • Freistaat Thüringen
  • European Union
  • Deutsche Forschungsgemeinschaft
  • German Research Foundation (DFG)
  • German Federal Ministry of Education and Research (BMBF)
  • Marie Skłodowska-Curie
  • EMBRACE

Zitieren

APA

Fiedler, P., Graichen, U., Zimmer, E., & Haueisen, J. (2023). Simultaneous Dry and Gel-Based High-Density Electroencephalography Recordings. Sensors. https://doi.org/10.3390/s23249745

Vancouver

Fiedler P, Graichen U, Zimmer E, Haueisen J. Simultaneous Dry and Gel-Based High-Density Electroencephalography Recordings. Sensors. 2023. doi:10.3390/s23249745

BibTeX
@article{Fiedler2023,
  author = {Fiedler, Patrique and Graichen, Uwe and Zimmer, Ellen and Haueisen, Jens},
  title = {Simultaneous Dry and Gel-Based High-Density Electroencephalography Recordings},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23249745}
}