Comparison of electrode position marking procedures on the cranial surface
Brain and Behavior · 2023
DOI: 10.1002/brb3.3187 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Brain and Behavior, 2023. Verfasst von Albert Fabregat-Sanjuan und 3 weiteren Autor:innen. Beteiligte Einrichtung: FUNCMAT, Mechanical Engineering Department, Universitat Rovira i Virgili, Tarragona, Spain. Hauptthemen laut MEDLINE: Skull. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitat Rovira i Virgili, "la Caixa" Foundation, European Observation Network for Territorial Development and Cohesion.
Abstract (Originalsprache)
Introduction
The study aimed to compare the conventional method of electrode marking with a new system, EPlacement, to improve accuracy and reduce the time burden on health care professionals.
Methods
Ten health care professionals marked mannequin heads and adult volunteers using both methods. Time, accuracy, and usability of each method were analyzed. Three neurophysiological diagnostic tests were performed on mannequin heads: reversal pattern visual evoked potential (three electrodes required); somatosensory evoked potentials from the upper and lower extremities (five electrodes required); and standard intraoperative neurophysiological monitoring for spine surgery (nine electrodes required). Precision scanning of the mannequins with structured light and a printed hull were used to determine the actual locations of the electrodes of the 10/20 system.
Results
The new method based on the EPlacement device represents an improvement on conventional tape measure (TM) marking and may be considered within the group of advanced methods such as navigation systems since it leads to improvements of 34% (1.7 mm) for electrode positions in the Nasion‐Inion and Left tragus–Right tragus lines and 77% (12.5 mm) for electrode positions using the approximate method. It reduces the time spent per test by an average of 1 min compared to the TM method. Health care staff survey results show a positive feedback regarding usability of the new method.
Conclusions
The study showed that the EPlacement device improves accuracy, reduces time, and is easy to use compared to the conventional method of electrode marking. The EPlacement method can facilitate the complex task of electrode marking and ultimately contribute to improved patient outcomes. It has the potential to be widely accepted and implemented in clinical practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- FUNCMAT, Mechanical Engineering Department, Universitat Rovira i Virgili, Tarragona, Spain.
- NeuroÈpia, Clinical Neurophysiology Department, Institut d'Investigació Sanitària Pere Virgili, Hospital Universitari Sant Joan de Reus, Tarragona, Spain.
- BIOMEC, Mechanical Engineering Department, Universitat Politècnica de Catalunya, Barcelona, Spain.
ORCID-Profile: Albert Fabregat-Sanjuan
Förderung
- Universitat Rovira i Virgili
- "la Caixa" Foundation
- European Observation Network for Territorial Development and Cohesion
Zitieren
APA
Fabregat-Sanjuan, A., Pàmies-Vilà, R., Rigo-Vidal, A., & Pascual-Rubio, V. (2023). Comparison of electrode position marking procedures on the cranial surface. Brain and Behavior. https://doi.org/10.1002/brb3.3187
Vancouver
Fabregat-Sanjuan A, Pàmies-Vilà R, Rigo-Vidal A, Pascual-Rubio V. Comparison of electrode position marking procedures on the cranial surface. Brain and Behavior. 2023. doi:10.1002/brb3.3187
BibTeX
@article{FabregatSanjuan2023,
author = {Fabregat-Sanjuan, Albert and Pàmies-Vilà, Rosa and Rigo-Vidal, Agnès and Pascual-Rubio, Vicenç},
title = {Comparison of electrode position marking procedures on the cranial surface},
journal = {Brain and Behavior},
year = {2023},
doi = {10.1002/brb3.3187}
}
EUVIMED