EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Monitoring of Cardiorespiratory Parameters during Sleep Using a Special Holder for the Accelerometer Sensor

Andrei Boiko iD, Maksym Gaiduk iD, Wilhelm Daniel Scherz iD, Andrea Gentili iD, Massimo Conti iD, Simone Orcioni iD, Natividad Martínez Madrid iD, Ralf Seepold iD

Sensors · 05.06.2023 · 23 (11) · 5351

DOI: 10.3390/s23115351 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 05.06.2023. Verfasst von Andrei Boiko und 7 weiteren Autor:innen. Beteiligte Einrichtung: Ubiquitous Computing Lab, Department of Computer Science, HTWG Konstanz-University of Applied Sciences, 78462 Konstanz, Germany. Hauptthemen laut MEDLINE: Signal Processing, Computer-Assisted, Sleep. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Carl Zeiss Foundation, MORPHEUS-Project “Non-invasive system for measuring parameters relevant to sleep quality”, HTWG Hochschule Konstanz University of Applied Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sleep is extremely important for physical and mental health. Although polysomnography is an established approach in sleep analysis, it is quite intrusive and expensive. Consequently, developing a non-invasive and non-intrusive home sleep monitoring system with minimal influence on patients, that can reliably and accurately measure cardiorespiratory parameters, is of great interest. The aim of this study is to validate a non-invasive and unobtrusive cardiorespiratory parameter monitoring system based on an accelerometer sensor. This system includes a special holder to install the system under the bed mattress. The additional aim is to determine the optimum relative system position (in relation to the subject) at which the most accurate and precise values of measured parameters could be achieved. The data were collected from 23 subjects (13 males and 10 females). The obtained ballistocardiogram signal was sequentially processed using a sixth-order Butterworth bandpass filter and a moving average filter. As a result, an average error (compared to reference values) of 2.24 beats per minute for heart rate and 1.52 breaths per minute for respiratory rate was achieved, regardless of the subject’s sleep position. For males and females, the errors were 2.28 bpm and 2.19 bpm for heart rate and 1.41 rpm and 1.30 rpm for respiratory rate. We determined that placing the sensor and system at chest level is the preferred configuration for cardiorespiratory measurement. Further studies of the system’s performance in larger groups of subjects are required, despite the promising results of the current tests in healthy subjects.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Signal Processing, Computer-AssistedSleepMaleFemaleHumansPolysomnographyRespiratory RateHeart RateAccelerometry

Stichwörter der Autor:innen

accelerometerballistocardiographycontactless measurementhealth monitoringheart raterespiratory ratesleep monitoringvital signals

Autor:innen und Institutionen

  • Ubiquitous Computing Lab, Department of Computer Science, HTWG Konstanz-University of Applied Sciences, 78462 Konstanz, Germany.
  • Dipartimento di Ingegneria dell'Informazione, Università Politecnica delle Marche, 60131 Ancona, Italy.
  • School of Informatics, Reutlingen University, 72762 Reutlingen, Germany.

ORCID-Profile: Andrei Boiko, Maksym Gaiduk, Wilhelm Daniel Scherz, Andrea Gentili, Massimo Conti, Simone Orcioni, Natividad Martínez Madrid, Ralf Seepold

Förderung

  • Carl Zeiss Foundation
  • MORPHEUS-Project “Non-invasive system for measuring parameters relevant to sleep quality”
  • HTWG Hochschule Konstanz University of Applied Sciences

Zitieren

APA

Boiko, A., Gaiduk, M., Scherz, W. D., Gentili, A., Conti, M., Orcioni, S., Martínez Madrid, N., & Seepold, R. (2023). Monitoring of Cardiorespiratory Parameters during Sleep Using a Special Holder for the Accelerometer Sensor. Sensors, 23(11), 5351. https://doi.org/10.3390/s23115351

Vancouver

Boiko A, Gaiduk M, Scherz WD, Gentili A, Conti M, Orcioni S, et al. Monitoring of Cardiorespiratory Parameters during Sleep Using a Special Holder for the Accelerometer Sensor. Sensors. 2023;23(11):5351. doi:10.3390/s23115351

BibTeX
@article{Boiko2023,
  author = {Boiko, Andrei and Gaiduk, Maksym and Scherz, Wilhelm Daniel and Gentili, Andrea and Conti, Massimo and Orcioni, Simone and Martínez Madrid, Natividad and Seepold, Ralf},
  title = {Monitoring of Cardiorespiratory Parameters during Sleep Using a Special Holder for the Accelerometer Sensor},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {11},
  pages = {5351},
  doi = {10.3390/s23115351}
}