Thermal Cameras for Overnight Measuring of Respiration in a Clinical Setting

Raquel Alves iD, Fokke van Meulen iD, Sebastiaan Overeem iD, Hennie Janssen iD, Pauline van Hirtum iD, Svitlana Zinger, Sander Stuijk iD

Sensors · 2025

DOI: 10.3390/s25195956 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Raquel Alves und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Eindhoven University of Technology, 5600MB Eindhoven, The Netherlands. Hauptthemen laut MEDLINE: Polysomnography, Respiration, Thermography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dutch Research Council, NWO/HTSM program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Thermal imaging is a non-contact method for monitoring respiration activity during sleep. In this study, we evaluated its clinical application during overnight recordings in a sleep clinic. Five thermal cameras were used to detect breaths, the estimated respiration rate (RR), and inter-breath intervals (IBIs) in seven adults undergoing diagnostic polysomnography (PSG). Forty-five minutes of recordings were selected, consisting of 12 motionless and event-free segments. The thermal videos were processed using an adapted pre-existing thermal video processing algorithm. The respiration signals generated with the thermal cameras were validated against simultaneously recorded signals from the PSG system, the current gold standard for monitoring sleep. The results show a mean absolute error (MAE) ranging between 0.64 and 0.91 breaths per minute for the RR. Breath detection showed a sensitivity of 96.3%, and a precision of 94.1%. The MAE obtained between IBIs was 0.48 s, and the mean IBI variability difference recorded was 3.9 percentage points. In addition, the results from this clinical study show that the use of all five cameras and a single camera revealed no statistically significant differences, demonstrating the work towards a robust system. This first study of thermal cameras for the assessment of respiration in a clinical setting shows us the potential application of thermal imaging in clinical practice for respiration monitoring and establishes a foundation for further implementation in assessing sleep-disordered breathing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PolysomnographyRespirationThermographyHumansMaleAdultFemaleRespiratory RateAlgorithmsMiddle AgedSleepSignal Processing, Computer-AssistedVideo RecordingMonitoring, Physiologic

Stichwörter der Autor:innen

polysomnographyremote thermographyrespiration ratesleepthermal cameras

Autor:innen und Institutionen

  • Department of Electrical Engineering, Eindhoven University of Technology, 5600MB Eindhoven, The Netherlands.
  • Centre for Sleep Medicine Kempenhaeghe, 5590AB Heeze, The Netherlands.

ORCID-Profile: Raquel Alves, Fokke van Meulen, Sebastiaan Overeem, Hennie Janssen, Pauline van Hirtum, Sander Stuijk

Förderung

  • Dutch Research Council
  • NWO/HTSM program

Zitieren

APA

Alves, R., van Meulen, F., Overeem, S., Janssen, H., van Hirtum, P., Zinger, S., & Stuijk, S. (2025). Thermal Cameras for Overnight Measuring of Respiration in a Clinical Setting. Sensors. https://doi.org/10.3390/s25195956

Vancouver

Alves R, van Meulen F, Overeem S, Janssen H, van Hirtum P, Zinger S, et al. Thermal Cameras for Overnight Measuring of Respiration in a Clinical Setting. Sensors. 2025. doi:10.3390/s25195956

BibTeX
@article{Alves2025,
  author = {Alves, Raquel and van Meulen, Fokke and Overeem, Sebastiaan and Janssen, Hennie and van Hirtum, Pauline and Zinger, Svitlana and Stuijk, Sander},
  title = {Thermal Cameras for Overnight Measuring of Respiration in a Clinical Setting},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25195956}
}
Kostenlos starten →