Finite Element Analysis of Collared Hip Prosthesis Cross-Sections Under Dynamic Loading and Wear Conditions for Durable Orthopedic Implant Design
Prosthesis · 03.04.2026 · 8 (4) · 36
DOI: 10.3390/prosthesis8040036 ↗Kurz gefasst
Fachartikel in der Zeitschrift Prosthesis, online erschienen am 03.04.2026. Verfasst von Chethan K N und 5 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objective: Traditional hip implant evaluations often overlook patient-specific dynamic loadings. This study investigates the performance of novel collared hip implant designs under walking conditions, focusing on geometric profiles and two common stem materials: Ti-6Al-4V and CoCr alloy.
Methods
Patient-specific dynamic forces were applied using commercial finite element analysis, adhering to ISO and ASTM standards. Four cross-sectional profiles—circular, elliptical, oval, and trapezoidal—were initially evaluated for induced stresses and displacements. Subsequently, wear characteristics at implant junctions were analyzed, comparing CoCr (MC 1) and Ti-6Al-4V (MC 2) stems. The study also assessed the impact of using Ultra-High Molecular Weight Polyethylene (UHMWPE) acetabular cups.
Results
The elliptical (CS 2) cross-sectional profile demonstrated superior performance. Junction analysis revealed that the CoCr stem (MC 1) exhibited a stem-to-head sliding distance four times higher and contact pressure 5.5 times higher than the Ti-6Al-4V stem (MC 2). Specifically, MC 1 showed 82% higher contact pressure and 89% greater sliding distance at the stem–head junction compared to MC 2. Additionally, utilizing UHMWPE cups effectively eliminated squeaking sounds attributed to CoCr cups due to superior wear resistance.
Conclusions
The combination of an elliptical (CS 2) cross-sectional profile with a Ti-6Al-4V stem and UHMWPE acetabular cup offers optimal performance. This configuration significantly reduces wear and contact pressure, suggesting enhanced functionality and durability for hip implants under dynamic loading conditions.
Zitieren
APA
N, C. K., Corda, J. V., Keni, L. G., Kalayarasan, M., Reginald, J., & Prathik, S. J. (2026). Finite Element Analysis of Collared Hip Prosthesis Cross-Sections Under Dynamic Loading and Wear Conditions for Durable Orthopedic Implant Design. Prosthesis, 8(4), 36. https://doi.org/10.3390/prosthesis8040036
Vancouver
N CK, Corda JV, Keni LG, Kalayarasan M, Reginald J, Prathik SJ. Finite Element Analysis of Collared Hip Prosthesis Cross-Sections Under Dynamic Loading and Wear Conditions for Durable Orthopedic Implant Design. Prosthesis. 2026;8(4):36. doi:10.3390/prosthesis8040036
BibTeX
@article{N2026,
author = {N, Chethan K and Corda, John Valerian and Keni, Laxmikant G. and Kalayarasan, M. and Reginald, Jonathan and Prathik, Sudhir Jain},
title = {Finite Element Analysis of Collared Hip Prosthesis Cross-Sections Under Dynamic Loading and Wear Conditions for Durable Orthopedic Implant Design},
journal = {Prosthesis},
year = {2026},
volume = {8},
number = {4},
pages = {36},
doi = {10.3390/prosthesis8040036}
}
EUVIMED