Fachartikel in der Zeitschrift Clinical Implant Dentistry and Related Research, online erschienen am 13.11.2026. Verfasst von Siyi Chen und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Implantology, School and Hospital of Stomatology, Cheeloo College of Medicine, Shandong University, Shandong Key Laboratory of Oral Dis…. Hauptthemen laut MEDLINE: Dental Prosthesis, Implant-Supported, Jaw, Edentulous, Mandible, Dental Prosthesis Design. Die Arbeit nennt 51 Literaturangaben. Gefördert von: Xi'an Detawell Software Technology Co. Ltd., Shandong Maier Medical Technology Co. Ltd., Shandong University..
Abstract (Originalsprache)
Objective
This study aimed to evaluate the longitudinal clinical outcomes and biomechanical characteristics of various mandibular fixed implant‐supported prostheses, focusing on cantilever presence, implant quantity, bone density, and framework materials.
Methods
A retrospective clinical analysis and finite element analysis (FEA) were synthesized. The clinical cohorts of 17 patients were grouped by prosthetic design (cantilever versus non‐cantilever) to assess cumulative survival rates (CSR), marginal bone loss (MBL), technical complication rates, and bone density loss (BDL) over a 3‐year follow‐up period. Concurrently, 32 models were simulated for four configurations: All‐on‐4 with or without cantilevers (C4/A4) and All‐on‐6 with or without cantilevers (C6/A6). Occlusal loading was simulated across varying bone densities and framework materials. The peri‐implant stress and strain were evaluated.
Results
Both clinical cohorts showed high 3‐year implant CSR. However, cantilever group exhibited significantly higher MBL, whereas the between‐group difference in BDL varied over time. FEA suggested that distal cantilevers amplified peri‐implant stress and strain. In cases of posterior bone deficiency, the C6 design yielded more favorable biomechanical performance than the A4 configuration. While Type I bone reduced strain in FEA, no significant association was detected between clinical bone type and either MBL or BDL. Zirconia framework more effectively minimized peri‐implant stress and strain than polymethyl methacrylate.
Conclusions
While both prosthetic designs yield acceptable short‐term survival rates, cantilever configurations showed higher simulated implant stress and peri‐implant bone strain, consistent with the greater MBL observed in the clinical cantilever group. In cases of posterior bone deficiency, a six‐implant design with a short cantilever may distribute stress more effectively than utilizing a highly tilted four‐implant configuration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Implantology, School and Hospital of Stomatology, Cheeloo College of Medicine, Shandong University, Shandong Key Laboratory of Oral Diseases, Shandong Engineering Research Center of Dental Materials and Or…
- Department of Prosthodontics, Shandong Provincial Hospital Affiliated to Shandong First Medical University, Jinan, Shandong, China.
ORCID-Profile: Siyi Chen, Yuqing Wang, Wenhao Zhao, Xiaoni Ma
Förderung
- Xi'an Detawell Software Technology Co. Ltd.
- Shandong Maier Medical Technology Co. Ltd.
- Shandong University.
- Shandong Provincial Natural Science Foundation
- Shandong Provincial Hospital Affiliated to Shandong First Medical University
- Natural Science Foundation of Shandong Province
Zitieren
APA
Chen, S., Wang, Y., Zhao, W., Chen, X., Zhou, F., Yu, W., & Ma, X. (2026). Clinical and Finite Element Evaluation of Fixed Implant Prosthesis Designs for the Edentulous Mandible. Clinical Implant Dentistry and Related Research, 28(6). https://doi.org/10.1111/cid.70194
Vancouver
Chen S, Wang Y, Zhao W, Chen X, Zhou F, Yu W, et al. Clinical and Finite Element Evaluation of Fixed Implant Prosthesis Designs for the Edentulous Mandible. Clinical Implant Dentistry and Related Research. 2026;28(6). doi:10.1111/cid.70194
BibTeX
@article{Chen2026,
author = {Chen, Siyi and Wang, Yuqing and Zhao, Wenhao and Chen, Xuejiao and Zhou, Fangyu and Yu, Wenqian and Ma, Xiaoni},
title = {Clinical and Finite Element Evaluation of Fixed Implant Prosthesis Designs for the Edentulous Mandible},
journal = {Clinical Implant Dentistry and Related Research},
year = {2026},
volume = {28},
number = {6},
doi = {10.1111/cid.70194}
}
EUVIMED