Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility

Juan Antonio Rodríguez-Rama, Leticia Presa Madrigal, Alfredo Marín Lázaro, Javier Maroto Lorenzo, Ana García Laso, Jorge L. Costafreda Mustelier, Domingo A. Martín-Sánchez

Metrology · 23.03.2026 · 6 (1) · 21

DOI: 10.3390/metrology6010021 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Metrology, online erschienen am 23.03.2026. Verfasst von Juan Antonio Rodríguez-Rama und 6 weiteren Autor:innen. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study presents the metrological validation of encapsulated DS18B20 digital temperature sensors. Eight units were tested, and seven were analysed (sensor 8 was excluded owing to a systematic failure). The evaluation was performed using a standard comparison calibration, where Tref was defined as the mean of two calibrated Pt-100 probes in a Julabo DYNEO DD 601F thermostatic bath, following the TH-001 procedure of the Spanish Centre of Metrology (CEM). Four validation tests were performed: Test 1 (E1, 20 to 75 °C), Test 2 (E2, 20 to 72 °C), and with an extended range, Test 3 (E3, −12 to 86 °C) and Test 4 (E4, −12 to 86 °C; repetition to assess reproducibility relative to E3), with 10 steady-state readings per setpoint. Erroneous readings were defined and removed (probe 3, Test 4), and set points without valid readings from probe 4 above 68 °C were excluded. Without data processing, the errors were consistent with the manufacturer’s stated ±0.5 °C, despite an inter-probe bias. Several correction models were evaluated (offset, affine linear, polynomial, and segmented); the probe-specific affine linear model provided the best overall compromise, reducing MAE (Mean Absolute Error) to 0.046 to 0.130 °C and RMSE (Root Mean Square Error) to 0.057 to 0.169 °C. The process uncertainty is dominated by the traceability of the Pt-100 probes and the effective nonuniformity of the isothermal volume, which limits the achievable accuracy. The results support the use of individually calibrated DS18B20 sensors for continuous monitoring, provided that the effective operating range is maintained.

Zitieren

APA

Rodríguez-Rama, J. A., Madrigal, L. P., Lázaro, A. M., Lorenzo, J. M., Laso, A. G., Mustelier, J. L. C., & Martín-Sánchez, D. A. (2026). Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility. Metrology, 6(1), 21. https://doi.org/10.3390/metrology6010021

Vancouver

Rodríguez-Rama JA, Madrigal LP, Lázaro AM, Lorenzo JM, Laso AG, Mustelier JLC, et al. Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility. Metrology. 2026;6(1):21. doi:10.3390/metrology6010021

BibTeX
@article{RodrguezRama2026,
  author = {Rodríguez-Rama, Juan Antonio and Madrigal, Leticia Presa and Lázaro, Alfredo Marín and Lorenzo, Javier Maroto and Laso, Ana García and Mustelier, Jorge L. Costafreda and Martín-Sánchez, Domingo A.},
  title = {Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and Reproducibility},
  journal = {Metrology},
  year = {2026},
  volume = {6},
  number = {1},
  pages = {21},
  doi = {10.3390/metrology6010021}
}
Kostenlos starten →