Microleakage and Torque Loss at the Implant–Abutment Interface in Original Versus Non-Original Abutments: An In Vitro Study
Materials · 02.05.2026 · 19 (9) · 1884
DOI: 10.3390/ma19091884 ↗Kurz gefasst
Fachartikel in der Zeitschrift Materials, online erschienen am 02.05.2026. Verfasst von Ferran Sánchez-Benito und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Stomatology, Faculty of Medicine and Dentistry, University of Valencia, 46010 Valencia, Spain. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Microleakage at the implant–abutment interface represents a potential pathway for bacterial penetration and may contribute to peri-implant inflammation, marginal bone loss, and mechanical complications such as screw loosening. The increasing clinical use of compatible prosthetic abutments as cost-effective alternatives to original components has raised concerns regarding their fit, sealing capacity, and mechanical stability at this interface. The aim of this in vitro study was to evaluate differences in sealing capacity and torque loss between original and non-original abutments in a mixed internal connection implant system and to investigate the applicability of a novel quantitative approach for assessing microleakage based on a hydraulic conductance perfusion system. Nine abutments, including four multi-unit and five screw-retained cementable abutments, were connected to Straumann Bone Level implants at two tightening torques (5 N·cm and 35 N·cm). Microleakage was quantified by measuring fluid transport across the implant–abutment interface using the perfusion system, and removal torque values were recorded after testing. Non-original abutments exhibited significantly greater microleakage than original abutments at both torque levels. Microleakage increased significantly when the installation torque was reduced to 5 N·cm. At the manufacturer-recommended torque, screw-retained cementable abutments demonstrated higher microleakage than multi-unit abutments. Non-original abutments also showed significantly greater torque loss. These findings suggest that original abutments provide improved sealing capacity and mechanical stability at the implant–abutment interface, while the hydraulic conductance perfusion system represents a promising quantitative tool for investigating microleakage.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Stomatology, Faculty of Medicine and Dentistry, University of Valencia, 46010 Valencia, Spain.
ORCID-Profile: Ferran Sánchez-Benito, Enrique Castells-Mira, María Cosin-Villanueva, Andrés López-Roldán
Zitieren
APA
Sánchez-Benito, F., Castells-Mira, E., Cosin-Villanueva, M., Gil-Loscos, F., & López-Roldán, A. (2026). Microleakage and Torque Loss at the Implant–Abutment Interface in Original Versus Non-Original Abutments: An In Vitro Study. Materials, 19(9), 1884. https://doi.org/10.3390/ma19091884
Vancouver
Sánchez-Benito F, Castells-Mira E, Cosin-Villanueva M, Gil-Loscos F, López-Roldán A. Microleakage and Torque Loss at the Implant–Abutment Interface in Original Versus Non-Original Abutments: An In Vitro Study. Materials. 2026;19(9):1884. doi:10.3390/ma19091884
BibTeX
@article{SnchezBenito2026,
author = {Sánchez-Benito, Ferran and Castells-Mira, Enrique and Cosin-Villanueva, María and Gil-Loscos, Francisco and López-Roldán, Andrés},
title = {Microleakage and Torque Loss at the Implant–Abutment Interface in Original Versus Non-Original Abutments: An In Vitro Study},
journal = {Materials},
year = {2026},
volume = {19},
number = {9},
pages = {1884},
doi = {10.3390/ma19091884}
}
EUVIMED