A Novel Tool for the Rapid and Transparent Verification of Reference Intervals in Clinical Laboratories
Journal of Clinical Medicine · 27.07.2024 · 13 (15) · 4397
DOI: 10.3390/jcm13154397 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 27.07.2024. Verfasst von Georg Hoffmann und 5 weiteren Autor:innen. Beteiligte Einrichtung: German Heart Center, Institute of Laboratory Medicine, Technical University Munich, 80636 Munich, Germany. Die Arbeit wurde bisher 32-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: We present a software package called reflimR (Version 1.0.6), which enables rapid and transparent verification of reference intervals from routine laboratory measurements. Our method makes it easy to compare the results with specified target values and facilitates the interpretation of deviations using traffic light colors.
Methods
The algorithm includes three procedural steps: (a) definition of an appropriate distribution model, based on Bowley’s quartile skewness, (b) iterative truncation, based on a modified boxplot method to obtain the central 95% of presumably inconspicuous results, and (c) extrapolation of reference limits from a truncated normal quantile–quantile plot.
Results
All algorithms have been combined into one consolidated library, which can be called in the R environment with a single command reflim (x). Using an example dataset included in the package, we demonstrate that our method can be applied to mixed data containing a substantial proportion of pathological values. It leads to similar results as the direct guideline approach as well as the more sophisticated indirect refineR software package. As compared to the latter, reflimR works much faster and needs smaller datasets for robust estimates. For the interpretation of the results, we present an intuitive color scheme based on tolerance ranges (permissible uncertainty of laboratory results). We show that a relatively high number of published reference limits require careful reevaluation.
Conclusions
The reflimR package closes the gap between direct guideline methods and the more sophisticated indirect refineR method. We recommend reflimR for the rapid routine verification of large amounts of reference limits and refineR for a careful analysis of unclear or doubtful results from this check.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- German Heart Center, Institute of Laboratory Medicine, Technical University Munich, 80636 Munich, Germany.
- Trillium GmbH Medizinischer Fachverlag, 82284 Grafrath, Germany.
- Institute for Information Engineering, Ostfalia University of Applied Sciences, 38302 Wolfenbüttel, Germany.
- Institut für Hämostaseologie und Pharmakologie (IHP), 12247 Berlin, Germany.
- Institut für Medizinische Diagnostik (IMD), 12247 Berlin, Germany.
- Helmholtz Centre for Infection Research, 38124 Braunschweig, Germany.
ORCID-Profile: Sandra Klawitter, Stefan Holdenrieder, Frank Klawonn
Zitieren
APA
Hoffmann, G., Klawitter, S., Trulson, I., Adler, J., Holdenrieder, S., & Klawonn, F. (2024). A Novel Tool for the Rapid and Transparent Verification of Reference Intervals in Clinical Laboratories. Journal of Clinical Medicine, 13(15), 4397. https://doi.org/10.3390/jcm13154397
Vancouver
Hoffmann G, Klawitter S, Trulson I, Adler J, Holdenrieder S, Klawonn F. A Novel Tool for the Rapid and Transparent Verification of Reference Intervals in Clinical Laboratories. Journal of Clinical Medicine. 2024;13(15):4397. doi:10.3390/jcm13154397
BibTeX
@article{Hoffmann2024,
author = {Hoffmann, Georg and Klawitter, Sandra and Trulson, Inga and Adler, Jakob and Holdenrieder, Stefan and Klawonn, Frank},
title = {A Novel Tool for the Rapid and Transparent Verification of Reference Intervals in Clinical Laboratories},
journal = {Journal of Clinical Medicine},
year = {2024},
volume = {13},
number = {15},
pages = {4397},
doi = {10.3390/jcm13154397}
}