Verification of harmonized reference intervals in Croatia
Biochemia medica · 15.06.2026 · 36 (2) · 311-322
DOI: 10.11613/bm.2026.020706 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biochemia medica, online erschienen am 15.06.2026. Verfasst von Iva Friščić und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Biochemistry and Laboratory Medicine, Merkur University Hospital, Zagreb, Croatia. Hauptthemen laut MEDLINE: Blood Chemical Analysis. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
The study of harmonized reference intervals (RIs) in Croatia was published in 2004. Since then, notable methodology and population characteristics changes have occurred in the Croatian population, highlighting the need for re-evaluation. This study aimed to verify harmonized RIs for hematology, coagulation, and clinical chemistry laboratory tests and assess their applicability in one clinical hospital center. Materials and methods: This verification study included 100 apparently healthy adults (40 males, 60 females; aged 22 to 75 years), selected using both a priori and a posteriori exclusion criteria following CLSI EP28-A3c guidelines. Blood samples were collected under standardized preanalytical conditions. Hematology analysis was performed on the Sysmex XN1000 analyzer (Sysmex, Kobe, Japan), coagulation testing on the Sysmex CS2500 analyzer (Siemens Healthcare Diagnostics Inc., Marburg, Germany), and clinical chemistry analysis on the Beckman Coulter DxC700AU (Beckman Coulter, Brea, USA) and Radiometer ABL90 FLEX (Radiometer, Brønshøj, Denmark) analyzers. Reference intervals were considered verified if at least 90% of results (18/20) fell within the predefined RIs.
Results
All tested hematology and coagulation RIs were successfully verified. Among 27 tested biochemical analytes, 25 RIs were verified in the first sample set. Total calcium and alkaline phosphatase RIs required additional verification. International harmonized RIs were adopted and successfully verified for both tests using two independent sample sets.
Conclusions
Most harmonized RIs in use in Croatia remain applicable to the adult population of one clinical hospital center. However, periodic re-evaluation is needed due to changes in analytical methods and population characteristics. Reference intervals from international harmonization projects can provide a valid alternative when local verification fails.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Biochemistry and Laboratory Medicine, Merkur University Hospital, Zagreb, Croatia.
- Department for Clinical Laboratory Diagnostics, Children's Hospital Srebrnjak, Zagreb, Croatia.
Zitieren
APA
Friščić, I., Burazer, I., Radeljak, A., Perkov, S., & Kardum Paro, M. M. (2026). Verification of harmonized reference intervals in Croatia. Biochemia medica, 36(2), 311-322. https://doi.org/10.11613/bm.2026.020706
Vancouver
Friščić I, Burazer I, Radeljak A, Perkov S, Kardum Paro MM. Verification of harmonized reference intervals in Croatia. Biochemia medica. 2026;36(2):311-322. doi:10.11613/bm.2026.020706
BibTeX
@article{Frii2026,
author = {Friščić, Iva and Burazer, Ida and Radeljak, Andrea and Perkov, Sonja and Kardum Paro, Mirjana Mariana},
title = {Verification of harmonized reference intervals in Croatia},
journal = {Biochemia medica},
year = {2026},
volume = {36},
number = {2},
pages = {311--322},
doi = {10.11613/bm.2026.020706}
}
EUVIMED