Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques

Cengiz Evli iD, Merve Önder, Ruben Pauwels iD, Mehmet Hakan Kurt, İsmail Doruk Koçyiğit iD, Gökhan Yazıcı, Kaan Orhan iD

Diagnostics · 15.12.2025 · 15 (24) · 3201

DOI: 10.3390/diagnostics15243201 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 15.12.2025. Verfasst von Cengiz Evli und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Ankara University, 06560 Ankara, Turkey. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purposes: Beam hardening artifacts caused by dental implants remain one of the most significant limitations of cone-beam computed tomography (CBCT), often compromising the evaluation of peri-implant bone and potentially masking critical diagnostic findings. Although metal artifact reduction (MAR) and noise-optimization filters such as the Adaptive Image Noise Optimizer (AINO) are widely available in commercial CBCT systems, their effectiveness varies depending on implant configuration and scanning parameters. A clearer understanding of how implant positioning influences artifact severity—together with how MAR and AINO perform under different conditions—is essential for improving diagnostic reliability. Materials and Methods: A fresh frozen cadaver head, with dental implants inserted using two configurations (C1 and C2), was scanned using different scan parameters, with and without metal artifact reduction and image optimization filters. The percentages of gray value alteration due to artifacts were evaluated, using registered pre-implant scans as a control. Regions of interest were defined by an experienced researcher. For the two implant conditions, ROIs were placed as follows: C1—lingual, buccal and mesial to the mesial implant; lingual, buccal and distal to the distal implant; and an additional ROI between the implants (n = 7); C2—lingual, buccal, mesial and distal to each implant (n = 8). For each ROI, the mean gray value was measured in five consecutive axial slices, and rescaled according to calibration points in air and soft tissue.

Results

Significant differences were found in gray values across configurations and scan modes. In the C2 configuration, combined MAR and AINO restored gray values in certain ROIs from 1.227 (OFF) to 1.223 (MAR+AINO), closely matching the control (1.227). In contrast, C1 showed limited improvement; for example, buccal ROI gray values decreased from 3.978 (OFF) to 3.323 (AINO) compared to the control (3.273), with no significant benefit from additional MAR.

Conclusion

Artifacts from implants can be significantly affected by their (relative) position and the use of MAR and AINO.

Schlagwörter

Stichwörter der Autor:innen

AINOCBCTMARartifactcadaver

Autor:innen und Institutionen

  • Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Ankara University, 06560 Ankara, Turkey.
  • Department of Dentistry and Oral Health, Aarhus University, 98000 Aarhus, Denmark.
  • Department of Oral and Maxillofacial Surgery, Kırıkkale University, 71450 Kırıkkale, Turkey.
  • Department of Physical Therapy and Rehabilitation, Faculty of Health Sciences, Gazi University, 06680 Ankara, Turkey.
  • School and Hospital of Stomatology, Cheeloo College of Medicine, Shandong University, Jinan 250012, China.

ORCID-Profile: Cengiz Evli, Ruben Pauwels, İsmail Doruk Koçyiğit, Kaan Orhan

Zitieren

APA

Evli, C., Önder, M., Pauwels, R., Kurt, M. H., Koçyiğit, İ. D., Yazıcı, G., & Orhan, K. (2025). Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques. Diagnostics, 15(24), 3201. https://doi.org/10.3390/diagnostics15243201

Vancouver

Evli C, Önder M, Pauwels R, Kurt MH, Koçyiğit İD, Yazıcı G, et al. Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques. Diagnostics. 2025;15(24):3201. doi:10.3390/diagnostics15243201

BibTeX
@article{Evli2025,
  author = {Evli, Cengiz and Önder, Merve and Pauwels, Ruben and Kurt, Mehmet Hakan and Koçyiğit, İsmail Doruk and Yazıcı, Gökhan and Orhan, Kaan},
  title = {Ex Vivo Quantitative Evaluation of Beam Hardening Artifacts at Various Implant Locations in Cone-Beam Computed Tomography Using Metal Artifact Reduction and Noise Reduction Techniques},
  journal = {Diagnostics},
  year = {2025},
  volume = {15},
  number = {24},
  pages = {3201},
  doi = {10.3390/diagnostics15243201}
}
Kostenlos starten →