Influence of Geometric Orifice Area on the Hemodynamic Performance of Prosthetic Aortic Valve Designs

Samikshya Neupane, Tarun Goswami

Designs · 07.11.2026 · 10 (6) · 127

DOI: 10.3390/designs10060127 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Designs, online erschienen am 07.11.2026. Verfasst von Samikshya Neupane und 1 weiteren Autor:innen. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Geometric orifice area (GOA) is a directly controllable design parameter in prosthetic heart valves because changes in leaflet configuration alter the available opening and the resulting flow field. This study comparatively evaluated three trileaflet prosthetic aortic valve designs with an identical 23 mm housing, strut arrangement, and annular architecture but different leaflet configurations, producing GOAs of 78.85, 147.96, and 156.29 mm2. Transient computational fluid dynamics simulations were performed in ANSYS Fluent 2025 R1 using Newtonian blood and the k-ω SST turbulence model. Three idealized pressure-loading combinations were evaluated using prescribed peak left-ventricular inlet/constant outlet pressures of 120/80, 140/90, and 160/100 mmHg. These conditions were treated as controlled loading states rather than complete physiological arterial pressure waveforms. Peak-systolic results were evaluated at t = 2.35 s. Increasing GOA from 78.85 to 156.29 mm2 increased forward mass flow by approximately 57.0% and reduced peak velocity by approximately 33.0% when averaged over the loading conditions, whereas mean wall shear stress changed by only −3.6%. Increasing the prescribed loading from 120/80 to 160/100 mmHg increased forward mass flow by 22.6%, peak velocity by 17.5%, mean wall shear stress by 34.9%, and mean vorticity by 21.1%. Normalized sensitivity analysis indicated that forward flow was more responsive to geometric variation, whereas wall shear stress was substantially more responsive to the prescribed loading condition. The response to GOA was also nonuniform across the sampled designs, indicating that incremental increases in opening area should not be assumed to produce proportional hemodynamic benefits. Within the investigated design and loading ranges, GOA exerted its strongest influence on flow delivery and jet velocity, whereas surface shear was more strongly influenced by the prescribed operating load. These findings support a design-oriented assessment in which the hemodynamic benefit of an increased opening is evaluated together with the specific leaflet geometry through which that opening is achieved.

Zitieren

APA

Neupane, S., & Goswami, T. (2026). Influence of Geometric Orifice Area on the Hemodynamic Performance of Prosthetic Aortic Valve Designs. Designs, 10(6), 127. https://doi.org/10.3390/designs10060127

Vancouver

Neupane S, Goswami T. Influence of Geometric Orifice Area on the Hemodynamic Performance of Prosthetic Aortic Valve Designs. Designs. 2026;10(6):127. doi:10.3390/designs10060127

BibTeX
@article{Neupane2026,
  author = {Neupane, Samikshya and Goswami, Tarun},
  title = {Influence of Geometric Orifice Area on the Hemodynamic Performance of Prosthetic Aortic Valve Designs},
  journal = {Designs},
  year = {2026},
  volume = {10},
  number = {6},
  pages = {127},
  doi = {10.3390/designs10060127}
}
Kostenlos starten →