Comparison of Prosthetic Valve Deformation and Hemodynamic Performance Between Bicuspid and Tricuspid Aortic Valve Anatomy After Transcatheter Aortic Valve Replacement
Journal of the American Heart Association · 12.11.2026
DOI: 10.1161/jaha.126.049690 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of the American Heart Association, online erschienen am 12.11.2026. Verfasst von Dejing Feng und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology National Center for Cardiovascular Disease, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical Col…. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
Background
This study compared prosthetic valve deformation and postprocedural hemodynamic performance between bicuspid and tricuspid aortic valve anatomy after transcatheter aortic valve replacement with the Venus‐A prosthesis.
Methods
We retrospectively evaluated patients who underwent transcatheter aortic valve replacement between 2012 and 2022 and underwent postoperative computed tomography within 1 year. Predischarge echocardiographic assessments included the mean transvalvular gradient and paravalvular leak. Computed tomography measurements included expansion ratio, ellipticity, implantation depth, asymmetric leaflet expansion, and hypoattenuated leaflet thickening. The leaflet inflow and valve waist were collectively defined as the functional valve portion.
Results
Among 212 patients (106 bicuspid and 106 tricuspid), the mean age was 74.2±7.6 years, and 43.9% were women. Patients with bicuspid anatomy had a significantly lower oversizing ratio and underwent balloon predilation and postdilation more frequently than those with tricuspid anatomy. Residual transvalvular gradients were higher with bicuspid than tricuspid anatomy (12.0 [interquartile range, 8.2–17.0] versus 10.0 [7.2–14.0] mm Hg; P =0.014), whereas moderate to severe paravalvular leak did not differ significantly. Prosthetic ellipticity was greater in bicuspid anatomy, but expansion ratios, including those within the functional valve portion, did not differ significantly. Implantation depth, asymmetric leaflet expansion, and hypoattenuated leaflet thickening incidence also did not differ significantly.
Conclusions
Following Venus‐A implantation, bicuspid anatomy was associated with greater prosthetic ellipticity and higher residual gradients, despite no significant difference in prosthetic expansion.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology National Center for Cardiovascular Disease, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College Beijing People's Republic of China.
ORCID-Profile: Zheng Zhou, Haowen Ma, Guannan Niu, Hongliang Zhang, Zhenyan Zhao, Erli Zhang, Bin Zhang, Yunqing Ye, Zhe Li, Moyang Wang, Yongjian Wu
Zitieren
APA
Feng, D., Zhou, Z., Ma, H., Niu, G., Zhang, H., Zhao, Z., Zhang, E., Zhang, B., Ye, Y., Li, Z., Wang, M., Zhao, J., & Wu, Y. (2026). Comparison of Prosthetic Valve Deformation and Hemodynamic Performance Between Bicuspid and Tricuspid Aortic Valve Anatomy After Transcatheter Aortic Valve Replacement. Journal of the American Heart Association. https://doi.org/10.1161/jaha.126.049690
Vancouver
Feng D, Zhou Z, Ma H, Niu G, Zhang H, Zhao Z, et al. Comparison of Prosthetic Valve Deformation and Hemodynamic Performance Between Bicuspid and Tricuspid Aortic Valve Anatomy After Transcatheter Aortic Valve Replacement. Journal of the American Heart Association. 2026. doi:10.1161/jaha.126.049690
BibTeX
@article{Feng2026,
author = {Feng, Dejing and Zhou, Zheng and Ma, Haowen and Niu, Guannan and Zhang, Hongliang and Zhao, Zhenyan and Zhang, Erli and Zhang, Bin and Ye, Yunqing and Li, Zhe and Wang, Moyang and Zhao, Jie and Wu, Yongjian},
title = {Comparison of Prosthetic Valve Deformation and Hemodynamic Performance Between Bicuspid and Tricuspid Aortic Valve Anatomy After Transcatheter Aortic Valve Replacement},
journal = {Journal of the American Heart Association},
year = {2026},
doi = {10.1161/jaha.126.049690}
}
EUVIMED