EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

A Portable, Foldable Negative-Pressure Aerosol-Containment System (FNPACS) for Aerosol Control During Aerosol-Generating Procedures

Bing Rui Huang, Fatimah Ibrahim iD, Ina Ismiarti Shariffuddin, Puteri Ainaa S Ibrahim, Li-Yen Chang iD, Karunan Joseph, Mas Sahidayana Mohktar iD, Noorjahan Haneem Md Hashim

Bioengineering · 2026-06-09 · 13 (6) · 669

DOI: 10.3390/bioengineering13060669 ↗

At a glance

Journal article in Bioengineering, published online on 2026-06-09. Written by Bing Rui Huang and 7 co-authors. Institution: Department of Biomedical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur 50603, Malaysia. 41 references. The full text is freely available (open access, CC BY 4.0). Funded by: Ministry of Science, Technology and Innovation, Ministry of Science and Technology.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Aerosol-generating procedures (AGPs) expose healthcare personnel to airborne pathogens and require portable engineering controls that can be integrated into routine clinical workflows. We developed a portable, foldable negative-pressure aerosol-containment system (FNPACS) combining adaptive fan control, an H14 high-efficiency particulate air (HEPA) filter, and a disposable metal-oxide prefilter in a mobile filtration module. Bench performance was evaluated using pressure-flow testing in accordance with National Environmental Balancing Bureau (NEBB) procedures and International Organization for Standardization (ISO) 14644-3, polyalphaolefin aerosol challenge testing, and smoke visualization, while an exploratory clinical study assessed environmental contamination via real-time reverse-transcription PCR (rRT-PCR) in 11 patients (31 assay analyses). Bench testing demonstrated HEPA filtration efficiencies of 99.994–99.997%, stable negative-pressure generation across fan duty cycles, no detectable downstream breakthrough beyond the HEPA filter under the tested conditions, and effective inward airflow on smoke testing. A Lagrangian discrete phase model (DPM) particle-tracking simulation further characterized size-dependent aerosol-surrogate transport. Under HEPA-ON active-extraction conditions, 73.0–86.1% of simulated 0.3–10 µm water-equivalent particles were transported to the HEPA suction pathway, while 13.9–27.0% were deposited on internal wall surfaces. In the clinical evaluation, SARS-CoV-2 RNA detection on environmental swabs was limited and predominantly low level. The clearest reproducible signal occurred on the top interior surface under HEPA-OFF conditions, whereas HEPA-ON detections were isolated or presumptive high-Ct signals without reproducible confirmation. These findings provide preliminary engineering and usability support for FNPACS as a feasible near-source aerosol-control platform for AGPs. The patient swab component should be interpreted as an exploratory, proof-of-concept assessment rather than confirmatory evidence of clinical containment efficiency because several clinical cases had non-supportive patient-related controls and were therefore not used in the primary containment interpretation.

Subjects

Author keywords

HEPA filtrationaerosol containmentaerosol-generating proceduresinfection controlnegative-pressure enclosureportable medical device

Authors and institutions

  • Department of Biomedical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur 50603, Malaysia.
  • Centre for Innovation in Medical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur 50603, Malaysia.
  • Department of Anaesthesiology, Faculty of Medicine, Universiti Malaya, Lembah Pantai, Kuala Lumpur 50603, Malaysia.
  • Department of Medical Microbiology, Faculty of Medicine, Universiti Malaya, Lembah Pantai, Kuala Lumpur 50603, Malaysia.

ORCID profiles: Fatimah Ibrahim, Li-Yen Chang, Mas Sahidayana Mohktar

Funding

  • Ministry of Science, Technology and Innovation
  • Ministry of Science and Technology

Cite

APA

Huang, B. R., Ibrahim, F., Shariffuddin, I. I., Ibrahim, P. A. S., Chang, L. Y., Joseph, K., Mohktar, M. S., & Hashim, N. H. M. (2026). A Portable, Foldable Negative-Pressure Aerosol-Containment System (FNPACS) for Aerosol Control During Aerosol-Generating Procedures. Bioengineering, 13(6), 669. https://doi.org/10.3390/bioengineering13060669

Vancouver

Huang BR, Ibrahim F, Shariffuddin II, Ibrahim PAS, Chang LY, Joseph K, et al. A Portable, Foldable Negative-Pressure Aerosol-Containment System (FNPACS) for Aerosol Control During Aerosol-Generating Procedures. Bioengineering. 2026;13(6):669. doi:10.3390/bioengineering13060669

BibTeX
@article{Huang2026,
  author = {Huang, Bing Rui and Ibrahim, Fatimah and Shariffuddin, Ina Ismiarti and Ibrahim, Puteri Ainaa S and Chang, Li-Yen and Joseph, Karunan and Mohktar, Mas Sahidayana and Hashim, Noorjahan Haneem Md},
  title = {A Portable, Foldable Negative-Pressure Aerosol-Containment System (FNPACS) for Aerosol Control During Aerosol-Generating Procedures},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {6},
  pages = {669},
  doi = {10.3390/bioengineering13060669}
}