Effect of Print Orientation and Post-Curing Duration on the Flexural Strength, Flexural Modulus and Vickers Microhardness of a 3D-Printed Occlusal Splint Resin
Bioengineering · 04.12.2025 · 12 (12) · 1322
DOI: 10.3390/bioengineering12121322 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 04.12.2025. Verfasst von Mustafa Sahin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthodontics, Faculty of Dentistry, Çukurova University, Adana 01250, Türkiye. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Additive manufacturing (AM) offers precision and efficiency in occlusal splint fabrication; however, the combined influence of build orientation and post-curing duration on the mechanical performance of splint resins remains insufficiently explored. This in vitro experimental study evaluated the effects of three build orientations (0°, 45°, and 90°) and three post-curing protocols (uncured, standard, and extended) on the flexural strength (FS), flexural modulus (FM) and Vickers hardness number (VHN) of a Class IIa biocompatible occlusal splint resin (NextDent Ortho Rigid). A total of 180 specimens were fabricated using a vat polymerization-type 3D printing system. Statistical analyses were conducted using one-way analyses of variance and Tukey’s tests at a significance level of α = 0.05. Both build orientation and post-curing duration significantly affected FS and VHN (p < 0.001). The combination of 45° build orientations and extended post-curing produced the highest FS (169.76 MPa) and FM (7502.17 MPa), exceeding values typically reported for 3D-printed splints, while the 90° orientation with extended curing achieved the highest VHN (21.88). Hardness gains, however, plateaued beyond standard curing, indicating a trade-off between strength and surface hardness. These results demonstrate that print orientation and post-curing time are decisive parameters in optimizing the mechanical performance of 3D-printed occlusal splints. For high-load clinical applications such as bruxism, prioritizing flexural strength over surface hardness may improve appliance longevity, supporting 45° orientation with extended curing as an evidence-based manufacturing approach.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthodontics, Faculty of Dentistry, Çukurova University, Adana 01250, Türkiye.
- Department of Orthodontics, Faculty of Dentistry, Ordu University, Ordu 52200, Türkiye.
- Department of Pedodontics, Faculty of Dentistry, Ordu University, Ordu 52200, Türkiye.
- Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Çukurova University, Adana 01250, Türkiye.
ORCID-Profile: Mustafa Sahin, Suleyman Kutalmış Buyuk, Huseyin Simsek, Tuncer Akdoğan, Safiyenur Bircan
Zitieren
APA
Sahin, M., Buyuk, S. K., Simsek, H., Akdoğan, T., Bircan, S., & Toroglu, M. S. (2025). Effect of Print Orientation and Post-Curing Duration on the Flexural Strength, Flexural Modulus and Vickers Microhardness of a 3D-Printed Occlusal Splint Resin. Bioengineering, 12(12), 1322. https://doi.org/10.3390/bioengineering12121322
Vancouver
Sahin M, Buyuk SK, Simsek H, Akdoğan T, Bircan S, Toroglu MS. Effect of Print Orientation and Post-Curing Duration on the Flexural Strength, Flexural Modulus and Vickers Microhardness of a 3D-Printed Occlusal Splint Resin. Bioengineering. 2025;12(12):1322. doi:10.3390/bioengineering12121322
BibTeX
@article{Sahin2025,
author = {Sahin, Mustafa and Buyuk, Suleyman Kutalmış and Simsek, Huseyin and Akdoğan, Tuncer and Bircan, Safiyenur and Toroglu, Mustafa Serdar},
title = {Effect of Print Orientation and Post-Curing Duration on the Flexural Strength, Flexural Modulus and Vickers Microhardness of a 3D-Printed Occlusal Splint Resin},
journal = {Bioengineering},
year = {2025},
volume = {12},
number = {12},
pages = {1322},
doi = {10.3390/bioengineering12121322}
}
EUVIMED