Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes
Frontiers in Medicine · 24.07.2025 · 12
DOI: 10.3389/fmed.2025.1605641 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 24.07.2025. Verfasst von Phillip T Yuhas und 4 weiteren Autor:innen. Beteiligte Einrichtung: College of Optometry, The Ohio State University, Columbus, OH, United States. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS.
Abstract (Originalsprache)
Introduction
This study evaluated the agreement between a third-generation (G3) ocular response analyzer (ORA) and a first-generation (G1) ORA, and tested the ability of the keratoconus match index (KMI) to identify keratoconus.
Methods
Healthy participants ( n = 149 eyes) and participants with keratoconus ( n = 78 eyes) were enrolled for this study. Four measurements were taken bilaterally using the G1 and G3 ORA. Goldmann-correlated intraocular pressure (IOPg), corneal-compensated IOP (IOPcc), corneal hysteresis (CH), waveform score, KMI, and waveform parameters area under the first applanation peak (p1area), area under the second applanation peak (p2area), width of the first applanation peak (w1), width of the second applanation peak (w2), height of the first applanation peak (h1), and height of the second applanation peak (h2) were recorded from the measurement with the highest waveform score in the left eye. Paired t -tests or Wilcoxon signed-rank tests were used to assess agreement between the devices, and receiver-operating characteristic curves determined the ability of KMI to identify eyes with keratoconus.
Results
There was no difference in IOPcc or IOPg between the devices in both cohorts. CH was significantly greater for the G3 than for the G1 in healthy participants but not in keratoconus participants. For both cohorts, measurements of waveform score, KMI, p1area, p2area, w2, h1, and h2 were greater for the G3 than for the G1. Only w1 was smaller for the G3 than for the G1. There was no difference in the ability of KMI to differentiate ectatic from healthy eyes between the devices.
Discussion
Although the G1 and G3 can identify keratoconus using KMI, there is meaningful variation between them in IOP and biomechanical outcome parameters. Thus, clinicians and researchers should compare results between the devices with caution and should state which generation produced the data.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- College of Optometry, The Ohio State University, Columbus, OH, United States.
- Department of Ophthalmology and Visual Sciences, College of Medicine, The Ohio State University Wexner Medical Center, Columbus, OH, United States.
- Department of Biomedical Engineering, College of Engineering, The Ohio State University, Columbus, OH, United States.
Förderung
- NEI NIH HHS
Zitieren
APA
Yuhas, P. T., Fortman, M. M., Nye, M., Mahmoud, A. M., & Roberts, C. J. (2025). Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes. Frontiers in Medicine, 12. https://doi.org/10.3389/fmed.2025.1605641
Vancouver
Yuhas PT, Fortman MM, Nye M, Mahmoud AM, Roberts CJ. Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes. Frontiers in Medicine. 2025;12. doi:10.3389/fmed.2025.1605641
BibTeX
@article{Yuhas2025,
author = {Yuhas, Phillip T and Fortman, Maddison M and Nye, Michael and Mahmoud, Ashraf M and Roberts, Cynthia J},
title = {Intraocular pressure and ocular biomechanical parameters vary between generations of the ocular response analyzer in healthy and ectatic eyes},
journal = {Frontiers in Medicine},
year = {2025},
volume = {12},
doi = {10.3389/fmed.2025.1605641}
}
EUVIMED