Evaluation of corneal biomechanical properties using the ocular response analyzer and the dynamic Scheimpflug-Analyzer Corvis ST in high pressure and normal pressure open-angle glaucoma patients
PLOS ONE · 26.01.2023 · 18 (1) · e0281017
DOI: 10.1371/journal.pone.0281017 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 26.01.2023. Verfasst von Lisa Ramm und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, University Hospital Carl Gustav Carus, TU Dresden, Dresden, Germany. Hauptthemen laut MEDLINE: Glaucoma, Open-Angle, Low Tension Glaucoma. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 80 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To characterize differences in corneal biomechanics in high (HPG) and normal pressure (NPG) primary open-angle glaucoma, and its association to disease severity.
Methods
Corneal biomechanical properties were measured using the Ocular Response Analyzer (ORA) and the dynamic Scheimpflug-Analyzer Corvis ST (CST). Disease severity was functionally assessed by automated perimetry (Humphrey field analyzer) and structurally with the Heidelberg Retina Tomograph. To avoid a possible falsification by intraocular pressure, central corneal thickness and age, which strongly influence ORA and CST measurements, group matching was performed. Linear mixed models and generalized estimating equations were used to consider inter-eye correlation.
Results
Following group matching, 60 eyes of 38 HPG and 103 eyes of 60 NPG patients were included. ORA measurement revealed a higher CRF in HPG than in NPG (P < 0.001). Additionally, the CST parameter integrated radius (P < 0.001) was significantly different between HPG and NPG. The parameter SSI (P < 0.001) representing corneal stiffness was higher in HPG than in NPG. Furthermore, regression analysis revealed associations between biomechanical parameters and indicators of disease severity. In HPG, SSI correlated to RNFL thickness. In NPG, dependencies between biomechanical readings and rim area, MD, and PSD were shown.
Conclusion
Significant differences in corneal biomechanical properties were detectable between HPG and NPG patients which might indicate different pathophysiological mechanisms underlying in both entities. Moreover, biomechanical parameters correlated to functional and structural indices of diseases severity. A reduced corneal deformation measured by dynamic methods was associated to advanced glaucomatous damage.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology, University Hospital Carl Gustav Carus, TU Dresden, Dresden, Germany.
- Faculty of Medicine Carl Gustav Carus, Department of Ophthalmology, TU Dresden, Dresden, Germany.
ORCID-Profile: Lisa Ramm, Robert Herber, Karin R Pillunat
Zitieren
APA
Ramm, L., Herber, R., Lorenz, G., Jasper, C. S., Pillunat, L. E., & Pillunat, K. R. (2023). Evaluation of corneal biomechanical properties using the ocular response analyzer and the dynamic Scheimpflug-Analyzer Corvis ST in high pressure and normal pressure open-angle glaucoma patients. PLOS ONE, 18(1), e0281017. https://doi.org/10.1371/journal.pone.0281017
Vancouver
Ramm L, Herber R, Lorenz G, Jasper CS, Pillunat LE, Pillunat KR. Evaluation of corneal biomechanical properties using the ocular response analyzer and the dynamic Scheimpflug-Analyzer Corvis ST in high pressure and normal pressure open-angle glaucoma patients. PLOS ONE. 2023;18(1):e0281017. doi:10.1371/journal.pone.0281017
BibTeX
@article{Ramm2023,
author = {Ramm, Lisa and Herber, Robert and Lorenz, Georg and Jasper, Carolin S and Pillunat, Lutz E and Pillunat, Karin R},
title = {Evaluation of corneal biomechanical properties using the ocular response analyzer and the dynamic Scheimpflug-Analyzer Corvis ST in high pressure and normal pressure open-angle glaucoma patients},
journal = {PLOS ONE},
year = {2023},
volume = {18},
number = {1},
pages = {e0281017},
doi = {10.1371/journal.pone.0281017}
}