EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Biocompatibility of 3D-printed vs. thermoformed and heat-cured intraoral appliances

Joanna Weżgowiec, Andrzej Małysa, Wojciech Szlasa, Julita Kulbacka, Agnieszka Chwiłkowska, Marek Ziętek, Mieszko Więckiewicz

Frontiers in Bioengineering and Biotechnology · 29.10.2024 · 12

DOI: 10.3389/fbioe.2024.1453888 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 29.10.2024. Verfasst von Joanna Weżgowiec und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Dentistry, Faculty of Dentistry, Wroclaw Medical University, Wroclaw, Poland. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Uniwersytet Medyczny im. Piastów Slaskich we Wroclawiu.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The development of additive manufacturing has the potential to revolutionize the fabrication of medical devices. This technology, also known as 3D printing, offers precise, cost-effective, and personalized approaches, which could be particularly beneficial in the production of intraoral appliances. Despite its promise, research on the biocompatibility of 3D-printed intraoral devices is still limited. Our study aims to address this gap.

Methods

We examined the cytotoxicity of materials processed via three techniques commonly used for the fabrication of different intraoral appliances: 3D printing (Dental LT Clear), thermoforming (Duran adjusted with Durasplint LC), and conventional heat-curing (Villacryl H Plus). We also investigated the impact of chemical or UVC disinfection on the biocompatibility of these materials. We assessed the biological effects induced in human gingival fibroblasts (HGFs) through both direct contact tests (MTT and LDH assays) and extract tests (PrestoBlue, DCF, and cell death type assays). Additionally, we observed changes in cellular morphology and migration rate under an inverted light microscope. The surface roughness of materials was evaluated using contact profilometry. Statistical analysis was conducted using two-way analysis of variance.

Results

Our findings suggest that all three fabrication techniques induced a slight cytotoxic effect in HGFs, as evidenced by both direct contact and extract tests. However, these materials could be considered nontoxic according to the ISO 10993-5:2009 norm, as the decrease in metabolic activity observed was always less than 30% compared to the untreated control.

Conclusion

This novel study confirms that 3D printing may be a safe alternative to conventional methods for fabricating intraoral appliances. However, further tests assessing the long-term intraoral usage are still needed.

Schlagwörter

Stichwörter der Autor:innen

additive manufacturingbiomedical and dental materialscytotoxicitydental resinsfibroblastsocclusal splintoral devicepolymer

Autor:innen und Institutionen

  • Department of Experimental Dentistry, Faculty of Dentistry, Wroclaw Medical University, Wroclaw, Poland.
  • Department of Molecular and Cellular Biology, Faculty of Pharmacy, Wroclaw Medical University, Wroclaw, Poland.
  • Department of Immunology and Bioelectrochemistry, State Research Institute Centre for Innovative Medicine, Vilnius, Lithuania.

Förderung

  • Uniwersytet Medyczny im. Piastów Slaskich we Wroclawiu

Zitieren

APA

Weżgowiec, J., Małysa, A., Szlasa, W., Kulbacka, J., Chwiłkowska, A., Ziętek, M., & Więckiewicz, M. (2024). Biocompatibility of 3D-printed vs. thermoformed and heat-cured intraoral appliances. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1453888

Vancouver

Weżgowiec J, Małysa A, Szlasa W, Kulbacka J, Chwiłkowska A, Ziętek M, et al. Biocompatibility of 3D-printed vs. thermoformed and heat-cured intraoral appliances. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1453888

BibTeX
@article{Wegowiec2024,
  author = {Weżgowiec, Joanna and Małysa, Andrzej and Szlasa, Wojciech and Kulbacka, Julita and Chwiłkowska, Agnieszka and Ziętek, Marek and Więckiewicz, Mieszko},
  title = {Biocompatibility of 3D-printed vs. thermoformed and heat-cured intraoral appliances},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2024},
  volume = {12},
  doi = {10.3389/fbioe.2024.1453888}
}
Kostenlos starten →