EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Displacement of synthetic cranioplasty implant due to mechanical failure of titanium fixation plating: illustrative case

Graham A Branscom iD, Evan M Ren, Michael B Lemonick, Bahie Ezzat iD, Ziad Rifi iD, Shahram Majidi iD, Christopher P Kellner iD, Tomoyoshi Shigematsu iD

Journal of Neurosurgery: Case Lessons · 01.06.2026 · 11 (22)

DOI: 10.3171/case2694 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neurosurgery: Case Lessons, erschienen am 01.06.2026. Verfasst von Graham A Branscom und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Icahn School of Medicine at Mount Sinai, New York, New York. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND Cranioplasty is a common procedure following decompressive craniectomy for traumatic brain injury or stroke, among other reasons, in order to repair a skull defect. Cranioplasty failure is an infrequent occurrence, but in the case of synthetic implants, it is usually caused by infection and incorrect sizing. However, trauma-induced displacement, breakage, and deformities of cranioplasty implants are less common. OBSERVATIONS The authors present a unique case of a cranioplasty failure due to displacement of a polyetheretherketone (PEEK) implant secondary to catastrophic breakage of the titanium fixation plates. The cause of displacement is unknown but is suspected to be trauma. A subsequent review of the literature contextualizes common reasons for failure among synthetic implants and suggests that fracture of synthetic implants or their fixation plates is extremely rare. LESSONS The displacement of a PEEK cranioplasty implant due to breakage of titanium fixation plates is a rare, albeit possible, complication of this neurosurgical procedure. Lifestyle modifications are paramount to reduce trauma risk and resultant cranioplasty failure. The addition of miniplates may also reduce implant displacement risk. Finally, since rates of failure differ by material, careful consideration of the type of material used for the implant should be incorporated into neurosurgical planning. https://thejns.org/doi/10.3171/CASE2694

Schlagwörter

Stichwörter der Autor:innen

PEEK implantcase reportcranioplasty failuretitanium

Autor:innen und Institutionen

  • Department of Neurosurgery, Icahn School of Medicine at Mount Sinai, New York, New York.

ORCID-Profile: Graham A Branscom, Bahie Ezzat, Ziad Rifi, Shahram Majidi, Christopher P Kellner, Tomoyoshi Shigematsu

Zitieren

APA

Branscom, G. A., Ren, E. M., Lemonick, M. B., Ezzat, B., Rifi, Z., Majidi, S., Kellner, C. P., & Shigematsu, T. (2026). Displacement of synthetic cranioplasty implant due to mechanical failure of titanium fixation plating: illustrative case. Journal of Neurosurgery: Case Lessons, 11(22). https://doi.org/10.3171/case2694

Vancouver

Branscom GA, Ren EM, Lemonick MB, Ezzat B, Rifi Z, Majidi S, et al. Displacement of synthetic cranioplasty implant due to mechanical failure of titanium fixation plating: illustrative case. Journal of Neurosurgery: Case Lessons. 2026;11(22). doi:10.3171/case2694

BibTeX
@article{Branscom2026,
  author = {Branscom, Graham A and Ren, Evan M and Lemonick, Michael B and Ezzat, Bahie and Rifi, Ziad and Majidi, Shahram and Kellner, Christopher P and Shigematsu, Tomoyoshi},
  title = {Displacement of synthetic cranioplasty implant due to mechanical failure of titanium fixation plating: illustrative case},
  journal = {Journal of Neurosurgery: Case Lessons},
  year = {2026},
  volume = {11},
  number = {22},
  doi = {10.3171/case2694}
}
Kostenlos starten →