Establishment of local diagnostic reference levels in mammography: multicenter retrospective study in Trujillo, Peru

Cecilia Querebalu Garcia, Paul Elias Panizo Olivos, Eduardo Garnique Millones, Christian Heyrner Barragan Mondragon, José Fernando Márquez Pachas, Eduardo Becquer Carrasco Solis

Revista Mexicana de Física · 01.07.2026 · 72 (4) · 1-8

DOI: 10.31349/revmexfis.72.041101 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Revista Mexicana de Física, online erschienen am 01.07.2026. Verfasst von Cecilia Querebalu Garcia und 5 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Radiation exposure in mammography can vary considerably across centers, potentially affecting both diagnostic quality and patient safety. In Trujillo, Peru, no locally adjusted absorbed dose values were previously available. Therefore, the objective of this study was to establish local diagnostic reference levels (LDRLs) for digital mammography within the EsSalud hospital network (Social Health Insurance) in Trujillo. A retrospective, multicenter study was conducted in three hospitals, analyzing 133 digital mammograms acquired with Hologic Selenia Dimensions systems. Technical parameters such as kilovoltage, tube load, compression force, compressed breast thickness, and average glandular dose (AGD) were collected and processed using automated systems. The LDRLs, defined as the 75th percentile of the AGD distribution, were determined for the craniocaudal (CC) and mediolateral oblique (MLO) projections. Multiple linear regression analysis was performed to identify independent predictors of AGD. The proposed LDRLs were 1.78 mGy for CC and 2.12 mGy for MLO, values well below the reference level of 3 mGy recommended by the Basic Safety Standards of the International Atomic Energy Agency (IAEA), and 30 − 39% lower than the regional LDRLs previously reported for direct digital radiography (DDR) systems in Latin America. Tube load (β = 0.886, p < 0.001 for CC; β = 0.572, p < 0.001 for MLO) and compressed breast thickness (β = 0.193, p < 0.001 for CC; β = 0.297, p = 0.001 for MLO) were identified as the most significant technical predictors of AGD. The established LDRLs demonstrate optimized dosimetric practices within the EsSalud network in Trujillo, ensuring that patient doses are appropriate, consistent, and aligned with international standards. Their implementation will facilitate ongoing dose monitoring, protocol standardization, and the strengthening of radiation protection in breast cancer screening programs.

Zitieren

APA

Garcia, C. Q., Olivos, P. E. P., Millones, E. G., Mondragon, C. H. B., Pachas, J. F. M., & Solis, E. B. C. (2026). Establishment of local diagnostic reference levels in mammography: multicenter retrospective study in Trujillo, Peru. Revista Mexicana de Física, 72(4), 1-8. https://doi.org/10.31349/revmexfis.72.041101

Vancouver

Garcia CQ, Olivos PEP, Millones EG, Mondragon CHB, Pachas JFM, Solis EBC. Establishment of local diagnostic reference levels in mammography: multicenter retrospective study in Trujillo, Peru. Revista Mexicana de Física. 2026;72(4):1-8. doi:10.31349/revmexfis.72.041101

BibTeX
@article{Garcia2026,
  author = {Garcia, Cecilia Querebalu and Olivos, Paul Elias Panizo and Millones, Eduardo Garnique and Mondragon, Christian Heyrner Barragan and Pachas, José Fernando Márquez and Solis, Eduardo Becquer Carrasco},
  title = {Establishment of local diagnostic reference levels in mammography: multicenter retrospective study in Trujillo, Peru},
  journal = {Revista Mexicana de Física},
  year = {2026},
  volume = {72},
  number = {4},
  pages = {1--8},
  doi = {10.31349/revmexfis.72.041101}
}
Kostenlos starten →