National diagnostic reference levels for digital diagnostic and screening mammography in Uganda

Denish Odongo iD, Alen Musisi, Richard Omara Okello, Felix Bongomin iD, Geoffrey Erem iD

PLOS ONE · 29.08.2024 · 19 (8) · e0294541

DOI: 10.1371/journal.pone.0294541 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 29.08.2024. Verfasst von Denish Odongo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, School of Medicine, Makerere University, Kampala, Uganda. Hauptthemen laut MEDLINE: Mammography, Breast Neoplasms. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Screening and diagnostic mammography are associated with some risk of radiation-induced breast cancer. This study was conducted to establish the National Diagnostic Reference Levels (NDRLs) for digital diagnostic and screening mammography in Uganda to achieve breast radiation dose optimization.

Methods

A cross-sectional study was conducted among female participants recruited by consecutive sampling from three selected hospitals with digital mammography in Uganda. The study variables extracted from the mammography machines were exposure factors, compressed breast thickness (CBT), and Average Glandular Dose (AGD) of two standard mammogram views. The stratified National DRL was derived by calculating the 75 th percentile of the AGD across all the samples at various CBT ranges for both screening and diagnostic mammography in craniocaudal (CC) and mediolateral oblique (MLO) views.

Results

We included 300 participants with mean ages of 50.28±9.32 and 47.45±13.45 years for the screening and diagnostic mammography, respectively. There were statistically significant positive correlations between AGD and exposure factors (mAs, kVp) (all p-values

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MammographyBreast NeoplasmsHumansFemaleUgandaMiddle AgedAdultCross-Sectional StudiesEarly Detection of CancerRadiation DosageReference ValuesMass ScreeningBreastAged

Autor:innen und Institutionen

  • Department of Radiology, School of Medicine, Makerere University, Kampala, Uganda.
  • Department of Microbiology and Immunology, Faculty of Medicine, Gulu University, Gulu, Uganda.
  • Department of Radiology, St Francis Hospital Nsambya, Kampala, Uganda.

ORCID-Profile: Denish Odongo, Felix Bongomin, Geoffrey Erem

Zitieren

APA

Odongo, D., Musisi, A., Okello, R. O., Bongomin, F., & Erem, G. (2024). National diagnostic reference levels for digital diagnostic and screening mammography in Uganda. PLOS ONE, 19(8), e0294541. https://doi.org/10.1371/journal.pone.0294541

Vancouver

Odongo D, Musisi A, Okello RO, Bongomin F, Erem G. National diagnostic reference levels for digital diagnostic and screening mammography in Uganda. PLOS ONE. 2024;19(8):e0294541. doi:10.1371/journal.pone.0294541

BibTeX
@article{Odongo2024,
  author = {Odongo, Denish and Musisi, Alen and Okello, Richard Omara and Bongomin, Felix and Erem, Geoffrey},
  title = {National diagnostic reference levels for digital diagnostic and screening mammography in Uganda},
  journal = {PLOS ONE},
  year = {2024},
  volume = {19},
  number = {8},
  pages = {e0294541},
  doi = {10.1371/journal.pone.0294541}
}