Total-Body Parametric Imaging Using Relative Patlak Plot
Journal of Nuclear Medicine · 2025
DOI: 10.2967/jnumed.124.268496 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine, 2025. Verfasst von Siqi Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California Davis Medical Center, Sacramento, California; and sqlli@ucdavis.edu. Hauptthemen laut MEDLINE: Whole Body Imaging, Positron-Emission Tomography, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
The standard Patlak plot, a simple yet efficient model, is widely used to describe irreversible tracer kinetics for dynamic PET imaging. Its widespread application to whole-body parametric imaging remains constrained because of the need for a full-time-course input function (e.g., 1 h). In this paper, we demonstrate the relative Patlak (RP) plot, which eliminates the need for the early-time input function, for total-body parametric imaging and its application to 20-min clinical scans acquired in list mode.
Methods
We conducted a theoretic analysis to indicate that the RP intercept b' is equivalent to a ratio of the SUV relative to the plasma concentration, whereas the RP slope Ki ' is equal to the standard Patlak Ki (net influx rate) multiplied by a global scaling factor for each subject. One challenge in applying RP to a short scan duration (e.g., 20 min) is the resulting high noise in the parametric images. We applied a self-supervised deep-kernel method for noise reduction. Using the standard Patlak plot as the reference, the RP method was evaluated for lesion quantification, lesion-to-background contrast, and myocardial visualization in total-body parametric imaging in 22 human subjects (12 healthy subjects and 10 cancer patients) who underwent a 1-h dynamic 18F-FDG scan. The RP method was also applied to the dynamic data reconstructed from a clinical standard 20-min list-mode scan either at 1 or 2 h after injection for 2 cancer patients.
Results
We demonstrated that it is feasible to obtain high-quality parametric images from 20-min scans using RP parametric imaging with a self-supervised deep-kernel noise-reduction strategy. The RP slope Ki ' was highly correlated with the standard Patlak Ki in lesions and major organs, demonstrating its quantitative potential across subjects. Compared with conventional SUVs, the Ki ' images significantly improved lesion contrast and enabled visualization of the myocardium for potential cardiac assessment. The application of the RP parametric imaging to the 2 clinical scans also showed similar benefits.
Conclusion
Using total-body PET with the RP approach, it is feasible to generate parametric images using data from a 20-min clinical list-mode scan.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of California Davis Medical Center, Sacramento, California; and sqlli@ucdavis.edu.
- Department of Radiology, University of California Davis Medical Center, Sacramento, California; and.
- Department of Biomedical Engineering, University of California at Davis, Davis, California.
Förderung
- NCI NIH HHS
- NIBIB NIH HHS
Zitieren
APA
Li, S., Abdelhafez, Y. G., Nardo, L., Cherry, S. R., Badawi, R. D., & Wang, G. (2025). Total-Body Parametric Imaging Using Relative Patlak Plot. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.124.268496
Vancouver
Li S, Abdelhafez YG, Nardo L, Cherry SR, Badawi RD, Wang G. Total-Body Parametric Imaging Using Relative Patlak Plot. Journal of Nuclear Medicine. 2025. doi:10.2967/jnumed.124.268496
BibTeX
@article{Li2025,
author = {Li, Siqi and Abdelhafez, Yasser G and Nardo, Lorenzo and Cherry, Simon R and Badawi, Ramsey D and Wang, Guobao},
title = {Total-Body Parametric Imaging Using Relative Patlak Plot},
journal = {Journal of Nuclear Medicine},
year = {2025},
doi = {10.2967/jnumed.124.268496}
}
EUVIMED