Acquisition Duration Optimization Using Visual Grading Regression in [ 68 Ga]FAPI-46 PET Imaging of Oncologic Patients
Journal of Nuclear Medicine Technology · 2024
DOI: 10.2967/jnmt.123.267156 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine Technology, 2024. Verfasst von Ted Nilsson und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Science, Intervention, and Technology, Karolinska Institutet, Stockholm, Sweden. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 29 Literaturangaben.
Abstract (Originalsprache)
Fibroblast activation protein is a promising target for oncologic molecular imaging with radiolabeled fibroblast activation protein inhibitors (FAPI) in a large variety of cancers. However, there are yet no published recommendations on how to set up an optimal imaging protocol for FAPI PET/CT. It is important to optimize the acquisition duration and strive toward an acquisition that is sufficiently short while simultaneously providing sufficient image quality to ensure a reliable diagnosis. The aim of this study was to evaluate the feasibility of reducing the acquisition duration of [68Ga]FAPI-46 imaging while maintaining satisfactory image quality, with certainty that the radiologist's ability to make a clinical diagnosis would not be affected. Methods: [68Ga]FAPI-46 PET/CT imaging was performed on 10 patients scheduled for surgical resection of suspected pancreatic cancer, 60 min after administration of 3.6 ± 0.2 MBq/kg. The acquisition time was 4 min/bed position, and the raw PET data were statistically truncated and reconstructed to represent images with an acquisition duration of 1, 2, and 3 min/bed position, additional to the reference images of 4 min/bed position. Four image quality criteria that focused on the ability to distinguish specific anatomic details, as well as perceived image noise and overall image quality, were scored on a 4-point Likert scale and analyzed with mixed-effects ordinal logistic regression.
Results
A trend toward increasing image quality scores with increasing acquisition duration was observed for all criteria. For the overall image quality, there was no significant difference between 3 and 4 min/bed position, whereas 1 and 2 min/bed position were rated significantly (P < 0.05) lower than 4 min/bed position. For the other criteria, all images with a reduced acquisition duration were rated significantly inferior to images obtained at 4 min/bed position.
Conclusion
The acquisition duration can be reduced from 4 to 3 min/bed position while maintaining satisfactory image quality. Reducing the acquisition duration to 2 min/bed position or lower is not recommended since it results in inferior-quality images so noisy that clinical interpretation is significantly disrupted.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Clinical Science, Intervention, and Technology, Karolinska Institutet, Stockholm, Sweden.
- Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, Huddinge, Sweden.
- Department of Biomedical Engineering and Health Systems, KTH Royal Institute of Technology, Huddinge, Sweden.
- Department of Upper Gastrointestinal Diseases, Karolinska University Hospital, Huddinge, Sweden.
- Department of Oncology-Pathology, Karolinska Institutet, Stockholm, Sweden.
- Department of Radiopharmacy, Karolinska University Hospital, Stockholm, Sweden; and.
Zitieren
APA
Nilsson, T., Rasinski, P., Smedby, Ö., Af Burén, S., Sparrelid, E., Löhr, J. M., Tran, T. A., Blomgren, A., Tzortzakakis, A., Axelsson, R., & Holstensson, M. (2024). Acquisition Duration Optimization Using Visual Grading Regression in [ 68 Ga]FAPI-46 PET Imaging of Oncologic Patients. Journal of Nuclear Medicine Technology. https://doi.org/10.2967/jnmt.123.267156
Vancouver
Nilsson T, Rasinski P, Smedby Ö, Af Burén S, Sparrelid E, Löhr JM, et al. Acquisition Duration Optimization Using Visual Grading Regression in [ 68 Ga]FAPI-46 PET Imaging of Oncologic Patients. Journal of Nuclear Medicine Technology. 2024. doi:10.2967/jnmt.123.267156
BibTeX
@article{Nilsson2024,
author = {Nilsson, Ted and Rasinski, Pawel and Smedby, Örjan and Af Burén, Siri and Sparrelid, Ernesto and Löhr, J Matthias and Tran, Thuy A and Blomgren, August and Tzortzakakis, Antonios and Axelsson, Rimma and Holstensson, Maria},
title = {Acquisition Duration Optimization Using Visual Grading Regression in [ 68 Ga]FAPI-46 PET Imaging of Oncologic Patients},
journal = {Journal of Nuclear Medicine Technology},
year = {2024},
doi = {10.2967/jnmt.123.267156}
}
EUVIMED