Optimization and impact of sensitivity mode on abbreviated scan protocols with population-based input function for parametric imaging of [18F]-FDG for a long axial FOV PET scanner

W Lan, H Sari, A Rominger, C la Fougère, F P Schmidt iD

European Journal of Nuclear Medicine and Molecular Imaging · 2024

DOI: 10.1007/s00259-024-06745-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Nuclear Medicine and Molecular Imaging, 2024. Verfasst von W Lan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, Tuebingen, Germany. Hauptthemen laut MEDLINE: Fluorodeoxyglucose F18. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Tübingen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The long axial field of view, combined with the high sensitivity of the Biograph Vision Quadra PET/CT scanner enables the precise deviation of an image derived input function (IDIF) required for parametric imaging. Traditionally, this requires an hour-long dynamic PET scan for [ 18 F]-FDG, which can be significantly reduced by using a population-based input function (PBIF). In this study, we expand these examinations and include the scanner’s ultra-high sensitivity (UHS) mode in comparison to the high sensitivity (HS) mode and evaluate the potential for further shortening of the scan time.

Methods

Patlak K i and DV estimates were determined by the indirect and direct Patlak methods using dynamic [ 18 F]-FDG data of 6 oncological patients with 26 lesions (0–65 min p.i.). Both sensitivity modes for different number/duration of PET data frames were compared, together with the potential of using abbreviated scan durations of 20, 15 and 10 min by using a PBIF. The differences in parametric images and tumour-to-background ratio (TBR) due to the shorter scans using the PBIF method and between the sensitivity modes were assessed.

Results

A difference of 3.4 ± 7.0% (K i ) and 1.2 ± 2.6% (DV) was found between both sensitivity modes using indirect Patlak and the full IDIF (0–65 min). For the abbreviated protocols and indirect Patlak, the UHS mode resulted in a lower bias and higher precision, e.g., 45–65 min p.i. 3.8 ± 4.4% (UHS) and 6.4 ± 8.9% (HS), allowing shorter scan protocols, e.g. 50–65 min p.i. 4.4 ± 11.2% (UHS) instead of 7.3 ± 20.0% (HS). The variation of K i and DV estimates for both Patlak methods was comparable, e.g., UHS mode 3.8 ± 4.4% and 2.7 ± 3.4% (K i ) and 14.4 ± 2.7% and 18.1 ± 7.5% (DV) for indirect and direct Patlak, respectively. Only a minor impact of the number of Patlak frames was observed for both sensitivity modes and Patlak methods. The TBR obtained with direct Patlak and PBIF was not affected by the sensitivity mode, was higher than that derived from the SUV image (6.2 ± 3.1) and degraded from 20.2 ± 12.0 (20 min) to 10.6 ± 5.4 (15 min). K i and DV estimate images showed good agreement (UHS mode, RC: 6.9 ± 2.3% (K i ), 0.1 ± 3.1% (DV), peak signal-to-noise ratio (PSNR): 64.5 ± 3.3 dB (K i ), 61.2 ± 10.6 dB (DV)) even for abbreviated scan protocols of 50–65 min p.i.

Conclusions

Both sensitivity modes provide comparable results for the full 65 min dynamic scans and abbreviated scans using the direct Patlak reconstruction method, with good K i and DV estimates for 15 min short scans. For the indirect Patlak approach the UHS mode improved the K i estimates for the abbreviated scans.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fluorodeoxyglucose F18HumansPositron Emission Tomography Computed TomographyImage Processing, Computer-AssistedMaleFemaleRadiopharmaceuticalsMiddle AgedTime FactorsAgedNeoplasmsPositron-Emission TomographySensitivity and Specificity

Stichwörter der Autor:innen

LAFOV PETParametric imagingPatlakPopulation-based input functionTotal-body PET

Substanzen

Fluorodeoxyglucose F18Radiopharmaceuticals

Autor:innen und Institutionen

  • Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, Tuebingen, Germany.
  • Advanced Clinical Imaging Technology, Siemens Healthineers International AG, Lausanne, Switzerland.
  • Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Cluster of Excellence iFIT (EXC 2180) "Image Guided and Functionally Instructed Tumor Therapies", University of Tuebingen, Tuebingen, Germany.
  • Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, Tuebingen, Germany. F.Schmidt@med.uni-tuebingen.de.
  • Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, University of Tuebingen, Tuebingen, Germany. F.Schmidt@med.uni-tuebingen.de.

ORCID-Profile: F P Schmidt

Förderung

  • Universitätsklinikum Tübingen

Zitieren

APA

Lan, W., Sari, H., Rominger, A., Fougère, C., & Schmidt, F. P. (2024). Optimization and impact of sensitivity mode on abbreviated scan protocols with population-based input function for parametric imaging of [18F]-FDG for a long axial FOV PET scanner. European Journal of Nuclear Medicine and Molecular Imaging. https://doi.org/10.1007/s00259-024-06745-3

Vancouver

Lan W, Sari H, Rominger A, Fougère C, Schmidt FP. Optimization and impact of sensitivity mode on abbreviated scan protocols with population-based input function for parametric imaging of [18F]-FDG for a long axial FOV PET scanner. European Journal of Nuclear Medicine and Molecular Imaging. 2024. doi:10.1007/s00259-024-06745-3

BibTeX
@article{Lan2024,
  author = {Lan, W and Sari, H and Rominger, A and Fougère, C la and Schmidt, F P},
  title = {Optimization and impact of sensitivity mode on abbreviated scan protocols with population-based input function for parametric imaging of [18F]-FDG for a long axial FOV PET scanner},
  journal = {European Journal of Nuclear Medicine and Molecular Imaging},
  year = {2024},
  doi = {10.1007/s00259-024-06745-3}
}
Kostenlos starten →