Manual Ventilation Performance With Safety Device in Normal Versus Decreased Lung Compliance: A Single-Center Simulation Study
Respiratory Care · 2025
DOI: 10.1089/respcare.12363 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Respiratory Care, 2025. Verfasst von Prasanna Kumar und 6 weiteren Autor:innen. Beteiligte Einrichtung: Drs. Kumar and Brady are affiliated with Alpert Medical School of Brown University, Department of Emergency Medicine, Providence, Rhode Island. Hauptthemen laut MEDLINE: Respiration, Artificial, Respiratory Therapy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben. Gefördert von: National Science Foundation.
Abstract (Originalsprache)
Background
Resuscitator bags are commonly utilized in acute care settings; however, poor performance occurs irrespective of a provider’s qualifications or experience. A new flow-limiting device (Sotair by SafeBVM, Boston, Massachusetts) limits inspiratory flow during manual ventilation, thus minimizing peak inspiratory pressures. This study examined the differences in flow, pressure, and tidal volume (V T ) during ventilation with a manual resuscitator connected to the flow-limiting device versus a mechanical ventilator.
Methods
Second-year respiratory therapy students were recruited from an advanced cardiovascular life support class. Participants conducted a 2-min trial of manually ventilating a test lung utilizing normal and decreased compliance settings with the flow-limiting device connected to an endotracheal tube. Demographic data on participants’ age were collected. The control group consisted of a mechanical ventilator providing ventilation with the same test lung and compliance settings. Mean differences were compared between the manual ventilation and control group.
Results
A total of 41 respiratory therapy students (71% female, 76% undergraduate) participated. The mean experience level using the bag-valve-mask was 6.71, and the mean confidence level was 8.02; the scale was 0–10 with high numbers indicating greater experience or confidence. A small but statistically significant difference was found in mean peak pressures between manual ventilation with the flow-limiting device (15 cm H 2 O) and the mechanical ventilator (13 cm H 2 O) for the normal lung setting ( P = .008) but not for the decreased compliance lung setting (23 cm H 2 O vs 23 cm H 2 O with the ventilator). There was a significant difference in mean V T between manual ventilation (412 mL) and the mechanical ventilator (460 mL) in the decreased compliance lung setting ( P = .003) but not the normal compliance setting (452 mL vs 474 mL with the ventilator).
Conclusions
Although there were some statistically significant differences between the 2 groups, these differences were not clinically important. Participants adequately manually ventilated with V T similar to a mechanical ventilator.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Drs. Kumar and Brady are affiliated with Alpert Medical School of Brown University, Department of Emergency Medicine, Providence, Rhode Island.
- Dr. Culbreth is affiliated with Georgia State University, Department of Respiratory Therapy Atlanta, Georgia; and American College of Medical Toxicology, Phoenix, Arizona.
- Drs. Gardenhire and Wu are affiliated with Georgia State University, Department of Respiratory Therapy, Atlanta, Georgia.
- Dr. Slutsky is affiliated with Li Ka Shing Knowledge Institute, St. Michael's Hospital, Unity Health Toronto, University of Toronto, Toronto, Canada.
- Dr. Kendall is affiliated with Alpert Medical School of Brown University, Department of Anesthesiology, Providence, Rhode Island.
Förderung
- National Science Foundation
Zitieren
APA
Kumar, P., Culbreth, R., Gardenhire, D. S., Slutsky, A. S., Wu, Y. J., Kendall, M. C., & Brady, M. F. (2025). Manual Ventilation Performance With Safety Device in Normal Versus Decreased Lung Compliance: A Single-Center Simulation Study. Respiratory Care. https://doi.org/10.1089/respcare.12363
Vancouver
Kumar P, Culbreth R, Gardenhire DS, Slutsky AS, Wu YJ, Kendall MC, et al. Manual Ventilation Performance With Safety Device in Normal Versus Decreased Lung Compliance: A Single-Center Simulation Study. Respiratory Care. 2025. doi:10.1089/respcare.12363
BibTeX
@article{Kumar2025,
author = {Kumar, Prasanna and Culbreth, Rachel and Gardenhire, Douglas S and Slutsky, Arthur S and Wu, Ying J and Kendall, Mark C and Brady, Mark F},
title = {Manual Ventilation Performance With Safety Device in Normal Versus Decreased Lung Compliance: A Single-Center Simulation Study},
journal = {Respiratory Care},
year = {2025},
doi = {10.1089/respcare.12363}
}
EUVIMED