Comparative Study of Automated End Tidal Control Versus Manual Fresh Gas Flow Adjustment with Respect to Gas Usage and Delivery during Low Flow Anesthesia
Anesthesia and Critical Care · 2020 · 02 (02)
DOI: 10.26502/acc.014 ↗Kurz gefasst
Fachartikel in der Zeitschrift Anesthesia and Critical Care, 2020. Verfasst von Adarshchandra Swami und 2 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref).
Abstract (Originalsprache)
Background
During low-flow manually-controlled anesthesia (MCA) the anesthetist keeps adjusting end-tidal oxygen (EtO2) and anesthetic concentrations (EtAA) constantly to ensure adequate and safe anesthesia. This study was performed to compare End Tidal Control and Manual Control of Fresh gas flow during low flow anesthesia.
Methods
Study was conducted on 84 ASA I and II patients 18-65 years age, of either sex, requiring general anesthesia for more than 2 hours. In MCA group (n = 42) target end-tidal desflurane (4%) and EtO2 concentrations (≥ 35%) were manually controlled by the anesthetist. In EtCA group (n = 42), same anesthetic machine with additional end-tidal control feature was used to reach same targets.
Results
The median desflurane consumption was 15.71 ml and 16.71 ml in ETCA and MCA respectively (p value =0.002) after one hour and 28.71ml and 32.98 ml in ETCA and MCA respectively (p value
Zitieren
APA
Swami, A., Arora, K., & Puppala, P. (2020). Comparative Study of Automated End Tidal Control Versus Manual Fresh Gas Flow Adjustment with Respect to Gas Usage and Delivery during Low Flow Anesthesia. Anesthesia and Critical Care, 02(02). https://doi.org/10.26502/acc.014
Vancouver
Swami A, Arora K, Puppala P. Comparative Study of Automated End Tidal Control Versus Manual Fresh Gas Flow Adjustment with Respect to Gas Usage and Delivery during Low Flow Anesthesia. Anesthesia and Critical Care. 2020;02(02). doi:10.26502/acc.014
BibTeX
@article{Swami2020,
author = {Swami, Adarshchandra and Arora, Kanika and Puppala, Padma},
title = {Comparative Study of Automated End Tidal Control Versus Manual Fresh Gas Flow Adjustment with Respect to Gas Usage and Delivery during Low Flow Anesthesia},
journal = {Anesthesia and Critical Care},
year = {2020},
volume = {02},
number = {02},
doi = {10.26502/acc.014}
}
EUVIMED