End-Tidal Control Versus Manual Control of Inhalational Anesthesia Delivery: A Randomized Controlled Noninferiority Trial
Anesthesia & Analgesia · 2024-10 · 139 (4) · 812-820
DOI: 10.1213/ane.0000000000007132 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Anesthesia & Analgesia, online erschienen am 19.07.2024. Verfasst von Melissa D McCabe und 4 weiteren Autor:innen. Beteiligte Einrichtung: From the Department of Anesthesiology, Loma Linda University School of Medicine, Loma Linda, California. Hauptthemen laut MEDLINE: Anesthesia, Inhalation, Anesthetics, Inhalation. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
BACKGROUND: Precise anesthesia delivery helps ensure amnesia, analgesia, and immobility. Conventionally, the end-tidal anesthetic concentration is maintained through manual adjustment of the fresh gas flow and anesthetic vaporizer output. Some anesthesia delivery systems can deliver and maintain clinician-selected end-tidal anesthetic agent (EtAA) concentration using a modified closed-loop system. We evaluated the performance of an End-tidal Control (EtC) system on the Aisys CS 2 anesthesia machine (GE HealthCare). We hypothesized EtC anesthetic delivery would be noninferior to manually controlled anesthetic delivery. METHODS: The Multi-site Anesthesia randomized controlled STudy of End-tidal control compared to conventional Results (MASTER) Trial evaluated anesthetic delivery in 210 adult patients receiving inhaled anesthesia. Patients were randomized to either EtC or manual control (MC) anesthetic delivery. The primary objective was to determine whether, compared to conventional anesthesia practice, EtC achieves and maintains clinician-specified EtAA and end-tidal oxygen (Et o 2 ) concentrations within defined noninferiority limits. Noninferiority was concluded if the lower limit of the 95% confidence interval (CI) of the difference between the percent duration within the acceptable range (5% of steady state or a margin of ~10% of each agent’s minimum alveolar concentration) for EtC and MC was ≥ −5% for both EtAA and Et o 2 . Secondary objectives included performance measures: response time: time required to attain 90% of the first desired EtAA, overshoot: amount the controller (or vaporizer delivery) exceeded the desired EtAA, and accuracy: average deviation from the desired EtAA. RESULTS: EtC achieved and sustained targeted EtAA and Et o 2 concentrations within the noninferiority threshold. The EtAA was within 5% of the desired value 98% ± 2.05% of the time with EtC compared to 45.7% ± 31.7% of the time with MC (difference 52.3% [95% CI, 45.9%–58.6%], P < .0001). For Et o 2 , EtC was within the noninferiority limit 86.3% ± 22.8% of the time compared with MC at 41% ± 33.3% ( P < .0001, difference 45.3% [95% CI, 36.1%–54.5%]). The median response time for achieving 90% of the initial EtAA desired value was 75 seconds with EtC and 158 seconds with MC ( P = .0013). EtC exhibited a median overshoot of 6.64% of the selected EtAA concentration, whereas MC often failed to reach the clinician’s desired value. The difference in median percent deviation from desired EtAA value was 15.7% ([95% CI, 13.5%–19.0%], P < 0001). CONCLUSIONS: EtC achieves and maintains the EtAA and Et o 2 concentration in a manner that is noninferior to manually controlled anesthesia delivery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- From the Department of Anesthesiology, Loma Linda University School of Medicine, Loma Linda, California.
- Department of Anesthesiology and Pediatrics, Duke University Medical Center, Durham, North Carolina.
- Department of Anesthesiology, Emory University School of Medicine, Atlanta, Georgia.
- Biostatistics Division of Clinical Operations Department, GE HealthCare, Chicago, Illinois.
- Department of Anesthesiology, Roy and Lucille Carver College of Medicine, University of Iowa Health Care, Iowa City, Iowa.
Zitieren
APA
McCabe, M. D., Dear, G. L., Klopman, M. A., Garg, K., & Seering, M. S. (2024). End-Tidal Control Versus Manual Control of Inhalational Anesthesia Delivery: A Randomized Controlled Noninferiority Trial. Anesthesia & Analgesia, 139(4), 812-820. https://doi.org/10.1213/ane.0000000000007132
Vancouver
McCabe MD, Dear GL, Klopman MA, Garg K, Seering MS. End-Tidal Control Versus Manual Control of Inhalational Anesthesia Delivery: A Randomized Controlled Noninferiority Trial. Anesthesia & Analgesia. 2024;139(4):812-820. doi:10.1213/ane.0000000000007132
BibTeX
@article{McCabe2024,
author = {McCabe, Melissa D and Dear, Guy de L and Klopman, Matthew A and Garg, Kritika and Seering, Melinda S},
title = {End-Tidal Control Versus Manual Control of Inhalational Anesthesia Delivery: A Randomized Controlled Noninferiority Trial},
journal = {Anesthesia & Analgesia},
year = {2024},
volume = {139},
number = {4},
pages = {812--820},
doi = {10.1213/ane.0000000000007132}
}
EUVIMED