Verification of the method for measuring erythrocyte filterability using the IDA-01 device

A. S. Suvorova, N. S. Kushnir, D. S. Prudinnik, E. A. Bovt, F. I. Ataullakhanov, Elena I. Sinauridze

Pediatric Hematology/Oncology and Immunopathology · 30.06.2026 · 25 (2) · 154-164

DOI: 10.24287/j.1137 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pediatric Hematology/Oncology and Immunopathology, online erschienen am 30.06.2026. Verfasst von A. S. Suvorova und 5 weiteren Autor:innen. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

A common cause of hereditary hemolytic anemia is mutations of erythrocyte membrane proteins, which lead to a reduced ability of red blood cells (RBCs) to deform and pass through narrow capillaries and interendothelial slits of the spleen. This shortens the lifespan of erythrocytes and may result in hemolytic anemia. Therefore, the ability of RBCs to pass through narrow capillaries may serve as a good prognostic marker of a shortened RBC lifespan due to premature RBC hemolysis. The measurement of RBC filterability through artificial filters with 3–5 μm pore size (which is slightly smaller than the diameter of RBCs) might be the most suitable diagnostic method. However, the use of this method is hampered by poor reproducibility of its previously described variants. A recently proposed modified method for measuring RBC filterability using the IDA-01 device has demonstrated good sensitivity and specificity in the diagnosis of hereditary spherocytosis. However, its analytical performance (accuracy and specificity) has not yet been studied. Aim: to evaluate the analytical performance of the erythrocyte filterability measuring method using the IDA-01 filtrometer. Materials and methods. In this study, we used RBCs from healthy donors isolated from venous blood and washed using a standard procedure. The following parameters were analyzed: repeatability of the results (mean relative error and coefficient of variation) when measuring buffer flow time through the same filter (n = 10); reproducibility of the donor erythrocyte filterability measurements in two independent series of experiments conducted at different sites on different days (n = 15 and n = 31); and the reference range for normal filterability. The biological sensitivity of the method was also measured using suspensions containing varying amounts of non-filterable cells (RBCs exposed to 0.08% glutaraldehyde). All statistical calculations were performed using OriginPro 8.1 (OriginLab Corp., USA). Results and discussion. The IDA-01 device demonstrated good repeatability of measurements (relative error 2.79%, coefficient of variation 3.5%); the detection limit was 0.2 s; the reproducibility of erythrocyte filterability measurements in the two experimental series did not differ significantly (Fisher's test); the mean filterability values in these series also did not differ significantly (modified Student's t-test). The method is not biased. The reference range of filterability for donors was 0.81–0.87 (mean ± 1 standard deviation). The detection limit for pathologically altered cells was 0.003%. The accuracy of the results depends only on correct sample preparation (the degree of leukocyte removal and a constant hematocrit level in erythrocyte suspension). When all necessary conditions are met, the device specificity approaches 100%.

Conclusion

The erythrocyte filterability test performed using the IDA-01 device has high sensitivity and specificity. It is very simple, does not require expensive equipment, and can be easily performed in any laboratory, making it a promising method for the diagnosis of hereditary spherocytosis.

Zitieren

APA

Suvorova, A. S., Kushnir, N. S., Prudinnik, D. S., Bovt, E. A., Ataullakhanov, F. I., & Sinauridze, E. I. (2026). Verification of the method for measuring erythrocyte filterability using the IDA-01 device. Pediatric Hematology/Oncology and Immunopathology, 25(2), 154-164. https://doi.org/10.24287/j.1137

Vancouver

Suvorova AS, Kushnir NS, Prudinnik DS, Bovt EA, Ataullakhanov FI, Sinauridze EI. Verification of the method for measuring erythrocyte filterability using the IDA-01 device. Pediatric Hematology/Oncology and Immunopathology. 2026;25(2):154-164. doi:10.24287/j.1137

BibTeX
@article{Suvorova2026,
  author = {Suvorova, A. S. and Kushnir, N. S. and Prudinnik, D. S. and Bovt, E. A. and Ataullakhanov, F. I. and Sinauridze, Elena I.},
  title = {Verification of the method for measuring erythrocyte filterability using the IDA-01 device},
  journal = {Pediatric Hematology/Oncology and Immunopathology},
  year = {2026},
  volume = {25},
  number = {2},
  pages = {154--164},
  doi = {10.24287/j.1137}
}