Fachartikel in der Zeitschrift Measurement Science and Technology, online erschienen am 02.05.2024. Verfasst von Nam Thanh Duong und 3 weiteren Autor:innen. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: Department of Science and Technology of Vietnam, Ministry of Natural Resources and Environment of Vietnam.
Abstract (Originalsprache)
As demand for particulate matter (PM) monitoring increases for scientific and regulatory purposes, the need for an accurate calibration process for these instruments will also increase. This article focuses on studying the factors that cause measurement uncertainty during the calibration process with the calibration facility (ManDust) for reference gravimetric instruments. Measurement uncertainty in this article included type A and type B uncertainties. Type A uncertainty mostly included repeatability and reproducibility while volume uncertainty predominately contributed to type B uncertainty. To clearly examine how these factors contributed to the combined uncertainty, the calibration process was carried out at a high concentration range from 0.4 to 1.6 mg m −3 . In general, the measurement uncertainty ranged from 1.95% to 6.65%. Further analysis showed that type A uncertainty contributed more to the total uncertainty than type B. In particular, uncertainty from volume calculation and measurement contributed over 80% of type B uncertainty. During the PM filter weighing process, the uncertainty during the conditioning process contributed the most to type B uncertainty. This article also conducted a sensitivity analysis and found that besides volume measurement, the filter weighing procedure was also sensitive to changes even though it only contributed less than 5% of type B uncertainty. Using high-quality devices and careful handling during these two procedures should be emphasized to reduce the uncertainty of the calibration process.
Förderung
- Department of Science and Technology of Vietnam, Ministry of Natural Resources and Environment of Vietnam
Zitieren
APA
Duong, N. T., Tran, T. S., Tran, H. T., & Nguyen, G. H. (2024). Analysis of measurement uncertainty for verification/calibration of particulate matter measuring instruments using the gravimetric method. Measurement Science and Technology, 35(8), 085801. https://doi.org/10.1088/1361-6501/ad41ff
Vancouver
Duong NT, Tran TS, Tran HT, Nguyen GH. Analysis of measurement uncertainty for verification/calibration of particulate matter measuring instruments using the gravimetric method. Measurement Science and Technology. 2024;35(8):085801. doi:10.1088/1361-6501/ad41ff
BibTeX
@article{Duong2024,
author = {Duong, Nam Thanh and Tran, Tung Son and Tran, Hoa Thi and Nguyen, Giang Hoang},
title = {Analysis of measurement uncertainty for verification/calibration of particulate matter measuring instruments using the gravimetric method},
journal = {Measurement Science and Technology},
year = {2024},
volume = {35},
number = {8},
pages = {085801},
doi = {10.1088/1361-6501/ad41ff}
}