Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method

Thomas Koshy iD, Yenny Lamazares, Susan Evans, Saeed Jortani iD

Diagnostics · 20.08.2026 · 16 (16) · 2648

DOI: 10.3390/diagnostics16162648 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 20.08.2026. Verfasst von Thomas Koshy und 3 weiteren Autor:innen. Beteiligte Einrichtung: Beeatus Solutions, Inc., Palatine, IL 60067, USA. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ALCOR Scientific, ALCOR Scientific, Smithfield, RI, USA.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Erythrocyte sedimentation rate (ESR) is one of the most widely used tests for assessing inflammatory conditions. Current guidelines require that Erythrocyte Sedimentation Rate (ESR) testing be performed within four hours of collection for samples stored at RT or within 24 h if refrigerated; this is challenging with dispersed collection sites. Thus, we investigated whether testing technology affects sample stability by comparing results obtained using iSED®, based on photometric rheology, with the classical Westergren method.

Methods

Samples were evaluated at multiple time intervals. ESR values from each timepoint were plotted against their Time 0 values, and correlation statistics were analyzed using Passing-Bablok regression analysis and a Stability Drift analysis. Samples were considered stable if the correlation statistics of the Passing-Bablok analysis were statistically indistinct from the identity line.

Results

Samples in the iSED/RT cohort were stable for up to 28 h after collection, while those in the iSED/4–8 °C cohort were stable for up to 48 h. Samples in the Westergren/RT cohort were stable for up to 10 h after collection, but significantly deteriorated at 12 h and beyond.

Conclusions

The iSED methodology provides accurate ESR results for samples stored up to 28 h at RT and 48 h at 4–8 °C, which was significantly superior to the Westergren method for which samples were stable at RT for only 10 h.

Schlagwörter

Stichwörter der Autor:innen

Westergren methodblood sample stabilityerythrocyte sedimentation rate (ESR)iSEDinflammationphotometric rheologyred blood cell aggregationrouleaux formation

Autor:innen und Institutionen

  • Beeatus Solutions, Inc., Palatine, IL 60067, USA.
  • Kentucky Clinical Trials Laboratory, Louisville, KY 40202, USA.
  • Biodecisions Consulting, Los Gatos, CA 95033, USA.
  • Department of Pathology and Laboratory Medicine, University of Louisville School of Medicine, Louisville, KY 40292, USA.

ORCID-Profile: Thomas Koshy, Saeed Jortani

Förderung

  • ALCOR Scientific
  • ALCOR Scientific, Smithfield, RI, USA

Zitieren

APA

Koshy, T., Lamazares, Y., Evans, S., & Jortani, S. (2026). Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method. Diagnostics, 16(16), 2648. https://doi.org/10.3390/diagnostics16162648

Vancouver

Koshy T, Lamazares Y, Evans S, Jortani S. Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method. Diagnostics. 2026;16(16):2648. doi:10.3390/diagnostics16162648

BibTeX
@article{Koshy2026,
  author = {Koshy, Thomas and Lamazares, Yenny and Evans, Susan and Jortani, Saeed},
  title = {Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {16},
  pages = {2648},
  doi = {10.3390/diagnostics16162648}
}
Kostenlos starten →