Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection
Biomedical Instrumentation & Technology · 2024
DOI: 10.2345/0899-8205-58.1.25 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Instrumentation & Technology, 2024. Verfasst von Krystina M Hopkins und 4 weiteren Autor:innen. Hauptthemen laut MEDLINE: Disinfection, Endoscopes. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
Borescope examinations of endoscope channels are commonly described in literature, but no studies on surgical instrument lumen inspection have been published recently. Inadequately processed surgical instruments have been implicated in patient infections. This study assessed the utility of borescopes for inspecting surgical instruments.
Methods
The study team inspected and photographed sterilized, patient-ready arthroscopic shaver handpieces and suction tips using a tablet camera and borescopes to characterize internal anatomy, defects found in lumens, and the impact of recleaning on debris or residues.
Results
Ten suctions and eight shavers were inspected. All suctions had internal ridges and suction holes that were perpendicular to the lumen. All shavers had visible ridges, elbows, and lever mechanisms inside lumens. Of the 18 instruments, 16 (88%) had internal features that appeared rough or jagged and 17 (94%) had visible debris or discoloration in the lumens. Recleaning efforts generally were effective for suctions, but multiple rounds of recleaning with enhanced steps were less effective for shavers, which were replaced. Researchers documented retained soil and brush bristles in several new shavers despite following manufacturer instructions for cleaning and found visible damage and discoloration within five uses.
Discussion
This study demonstrated the value of borescope examinations for surgical instrument lumens. Visual inspections identified anatomical features that could influence cleaning effectiveness and detected residual soil, discoloration, and debris in most instruments. The findings suggested that manufacturer cleaning instructions were insufficient and additional cleaning was not always effective. In response, the site's multidisciplinary team strengthened risk assessment protocols and enhanced their cleaning practices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Zitieren
APA
Hopkins, K. M., Adams, S. J., Lamb, L. A., Smart, A. G., & Ofstead, C. L. (2024). Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection. Biomedical Instrumentation & Technology. https://doi.org/10.2345/0899-8205-58.1.25
Vancouver
Hopkins KM, Adams SJ, Lamb LA, Smart AG, Ofstead CL. Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection. Biomedical Instrumentation & Technology. 2024. doi:10.2345/0899-8205-58.1.25
BibTeX
@article{Hopkins2024,
author = {Hopkins, Krystina M and Adams, Steven J and Lamb, Larry A and Smart, Abigail G and Ofstead, Cori L},
title = {Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection},
journal = {Biomedical Instrumentation & Technology},
year = {2024},
doi = {10.2345/0899-8205-58.1.25}
}
EUVIMED