Effect of an automated flexible endoscope channel brushing system on improving reprocessing quality: a randomized controlled study
Endoscopy · 01.07.2023 · 55 (07) · 636-642
DOI: 10.1055/a-2009-4735 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Endoscopy, erschienen am 01.07.2023. Verfasst von Renduo Shang und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology, Renmin Hospital of Wuhan University, Wuhan, China. Hauptthemen laut MEDLINE: Endoscopes, Colonoscopes. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 19 Literaturangaben. Gefördert von: Innovation Team Project of Health Commission of Hubei Province.
Abstract (Originalsprache)
Background
Qualified reprocessing, of which meticulous channel brushing is the most crucial step, is essential for prevention and control of endoscopy-associated infections. However, channel brushing is often omitted in practice. This study aimed to evaluate the effect of an automated flexible endoscope channel brushing system (AECBS) on improving the quality of endoscope reprocessing.
Methods
This prospective, randomized controlled study was conducted between 24 November 2021 and 22 January 2022 at Renmin Hospital of Wuhan University, China. Eligible endoscopes were randomly allocated to the auto group (channels brushed by AECBS) or the manual group (channels brushed manually), with sampling and culturing after high-level disinfection and drying. The primary end point was the proportion of endoscopes with positive cultures.
Results
204 endoscopes in the auto group and 205 in the manual group were analyzed. The proportion of endoscopes with positive cultures was significantly lower in the auto group (15.2 % [95 %CI 10.7 %–21.0 %]) than in the manual group (23.4 % [95 %CI 17.9 %–29.9 %]).
Conclusions
AECBS could effectively reduce bioburden and improve reprocessing quality of gastroscopes and colonoscopes. AECBS has the potential to replace manual brushing and lower the risk of endoscopy-associated infections, providing a new option for the optimization of reprocessing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Gastroenterology, Renmin Hospital of Wuhan University, Wuhan, China.
- Key Laboratory of Hubei Province for Digestive System Disease, Renmin Hospital of Wuhan University, Wuhan, China.
- Hubei Provincial Clinical Research Center for Digestive Disease Minimally Invasive Incision, Renmin Hospital of Wuhan University, Wuhan, China.
- Department of Nursing, Renmin Hospital of Wuhan University, Wuhan, China.
- Department of Laboratory Science, Renmin Hospital of Wuhan University, Wuhan, China.
Förderung
- Innovation Team Project of Health Commission of Hubei Province
Zitieren
APA
Shang, R., Liu, J., Luo, Z., Huang, X., Zhang, C., Chen, D., Wu, L., Yao, L., Wang, X., Wang, Q., Wu, Y., Zhang, L., & Yu, H. (2023). Effect of an automated flexible endoscope channel brushing system on improving reprocessing quality: a randomized controlled study. Endoscopy, 55(07), 636-642. https://doi.org/10.1055/a-2009-4735
Vancouver
Shang R, Liu J, Luo Z, Huang X, Zhang C, Chen D, et al. Effect of an automated flexible endoscope channel brushing system on improving reprocessing quality: a randomized controlled study. Endoscopy. 2023;55(07):636-642. doi:10.1055/a-2009-4735
BibTeX
@article{Shang2023,
author = {Shang, Renduo and Liu, Jun and Luo, Zi and Huang, Xingmin and Zhang, Caiqing and Chen, Di and Wu, Lianlian and Yao, Liwen and Wang, Xianglan and Wang, Qing and Wu, Yunxing and Zhang, Lili and Yu, Honggang},
title = {Effect of an automated flexible endoscope channel brushing system on improving reprocessing quality: a randomized controlled study},
journal = {Endoscopy},
year = {2023},
volume = {55},
number = {07},
pages = {636--642},
doi = {10.1055/a-2009-4735}
}
EUVIMED