Storage cabinet design and infection prevention implications for long endoscopes
American Journal of Infection Control · 2025
DOI: 10.1016/j.ajic.2025.08.030 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Infection Control, 2025. Verfasst von Daisuke Ohki und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Infection Control and Prevention, The University of Tokyo Hospital, Tokyo, Japan; Department of Gastroenterology, Graduate School of Me…. Hauptthemen laut MEDLINE: Endoscopes, Infection Control, Desiccation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Nosokomiale Infektionen
Abstract (Originalsprache)
BACKGROUND: Adequate drying of endoscope channels following reprocessing is critical to inhibit microbial growth. This study evaluated residual moisture and bacterial contamination in endoscopes stored using different cabinets and storage methods. METHODS: EN-580T endoscopes were stored in 3 cabinets: TJ-908S and TJ-804S (both with active drying) and TM-804S (passive drying). Storage methods included full vertical suspension, partial suspension, and looped storage. Residual moisture was measured at 24 and 48 hours. Endoscopes were also inoculated with Pseudomonas aeruginosa (10⁵ CFU/mL) and stored under each condition. Bacterial cultures were performed to assess microbial growth. RESULTS: In TJ-908S with full vertical suspension, average residual moisture was 23.9 μL at 24 hours and 17.6 μL at 48 hours. In TM-804S and TJ-804S using partial suspension or looped storage, moisture ranged from 289.4 to 383.7 μL at 24 hours and from 278.5 to 518.7 μL at 48 hours. Bacterial cultures from TJ-908S-stored scopes showed minimal growth. In contrast, those stored in TM-804S and TJ-804S exhibited variable but higher CFU counts, including samples with > 100 CFU. CONCLUSIONS: Cabinet design and storage method significantly influence endoscope drying and bacterial proliferation. Full vertical suspension may reduce contamination risk and enhance infection control practices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Infection Control and Prevention, The University of Tokyo Hospital, Tokyo, Japan; Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
- Department of Infection Control and Prevention, The University of Tokyo Hospital, Tokyo, Japan.
- Department of Clinical Engineering, The University of Tokyo Hospital, Tokyo, Japan.
- Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
- Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan; Next-Generation Endoscopic Computer Vision, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
- Department of Infection Control and Prevention, The University of Tokyo Hospital, Tokyo, Japan. Electronic address: takeyatsutsumi@g.ecc.u-tokyo.ac.jp.
Zitieren
APA
Ohki, D., Mizoguchi, M., Iwazaki, K., Nakagawa, N., Tsuboi, M., Yoneda, R., Kitaura, S., Okushin, K., Satoh, N., Kakushima, N., Tsuji, Y., Fujishiro, M., & Tsutsumi, T. (2025). Storage cabinet design and infection prevention implications for long endoscopes. American Journal of Infection Control. https://doi.org/10.1016/j.ajic.2025.08.030
Vancouver
Ohki D, Mizoguchi M, Iwazaki K, Nakagawa N, Tsuboi M, Yoneda R, et al. Storage cabinet design and infection prevention implications for long endoscopes. American Journal of Infection Control. 2025. doi:10.1016/j.ajic.2025.08.030
BibTeX
@article{Ohki2025,
author = {Ohki, Daisuke and Mizoguchi, Miyuki and Iwazaki, Keigo and Nakagawa, Norio and Tsuboi, Mayo and Yoneda, Ryu and Kitaura, Satoshi and Okushin, Kazuya and Satoh, Nobuhiko and Kakushima, Naomi and Tsuji, Yosuke and Fujishiro, Mitsuhiro and Tsutsumi, Takeya},
title = {Storage cabinet design and infection prevention implications for long endoscopes},
journal = {American Journal of Infection Control},
year = {2025},
doi = {10.1016/j.ajic.2025.08.030}
}
EUVIMED