EUVIMEDEUVIMED entdecken
Crossref

Evaluation of slot height accuracy, symmetry, and slot wall parallelism in 3D-printed orthodontic brackets, and conventional ceramic brackets

Dylan Handlin, Gabriel Miranda, Michael Lee, Julie McCray, Patrick Foley, Ki Beom Kim

The Angle Orthodontist · 24.03.2026

DOI: 10.2319/040325-269.1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Angle Orthodontist, erschienen am 24.03.2026. Verfasst von Dylan Handlin und 5 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate and compare the slot height accuracy, symmetry, and slot wall parallelism among esthetic three-dimensional (3D) printed orthodontic brackets and conventional esthetic ceramic brackets. Materials and Methods An in vitro assessment was conducted using 240 esthetic orthodontic brackets (80 per group) with nominal 0.022-inch slot heights. Slot dimensions at the base and face, symmetry, and slot wall parallelism were analyzed through magnified images captured with a Canon DSLR camera and measured utilizing ImageJ software. Accuracy of slot height was evaluated against the tolerance limits (DIN 13971-2).

Results

LightForce brackets demonstrated superior accuracy, achieving slot heights closest to nominal dimensions, with high symmetry and minimal divergence in wall parallelism. Braces on Demand brackets exhibited slightly oversized slot dimensions compared to the nominal value and some asymmetry. In contrast, 3M Clarity brackets displayed the most significant deviations, including oversized slot heights, marked asymmetry, and pronounced slot wall divergence.

Conclusions

This study underscores the potential of 3D printing technology to enhance bracket fabrication precision over conventional ceramic options. These findings advocate further exploration of clinical applications of 3D printing to optimize manufacturing processes and treatment outcomes.

Zitieren

APA

Handlin, D., Miranda, G., Lee, M., McCray, J., Foley, P., & Kim, K. B. (2026). Evaluation of slot height accuracy, symmetry, and slot wall parallelism in 3D-printed orthodontic brackets, and conventional ceramic brackets. The Angle Orthodontist. https://doi.org/10.2319/040325-269.1

Vancouver

Handlin D, Miranda G, Lee M, McCray J, Foley P, Kim KB. Evaluation of slot height accuracy, symmetry, and slot wall parallelism in 3D-printed orthodontic brackets, and conventional ceramic brackets. The Angle Orthodontist. 2026. doi:10.2319/040325-269.1

BibTeX
@article{Handlin2026,
  author = {Handlin, Dylan and Miranda, Gabriel and Lee, Michael and McCray, Julie and Foley, Patrick and Kim, Ki Beom},
  title = {Evaluation of slot height accuracy, symmetry, and slot wall parallelism in 3D-printed orthodontic brackets, and conventional ceramic brackets},
  journal = {The Angle Orthodontist},
  year = {2026},
  doi = {10.2319/040325-269.1}
}
Kostenlos starten →