Heart Rate Estimation Using the Galaxy Watch During Maximal Cardiopulmonary Exercise Testing: Cross-Sectional Validation Study
JMIR Cardio · 16.04.2026 · 10 · e81917-e81917
DOI: 10.2196/81917 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift JMIR Cardio, online erschienen am 16.04.2026. Verfasst von Allan Inoue und 6 weiteren Autor:innen. Beteiligte Einrichtung: Health Data Lab, Sidia Institute of Science and Technology, Av. Autaz Mirim, 2211 - Distrito Industrial I, Manaus, 69075-155, Brazil, 55 (92) 3212-34…. Hauptthemen laut MEDLINE: Exercise Test, Heart Rate, Photoplethysmography. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Photoplethysmography-based smartwatches are increasingly used for continuous heart rate (HR) monitoring. Their accuracy has been demonstrated at rest or during low-intensity activity, but data are scarce for maximal-intensity exercise, when motion artifacts and rapid hemodynamic changes can degrade the photoplethysmography signal. Validating these devices under such demanding conditions is essential before they are applied to clinical exercise testing, athletic training, or remote health monitoring.
Objective
This study aimed to evaluate the validity of the Samsung Galaxy Watch 6 (GW6) in estimating HR throughout a graded, maximal ramp cardiopulmonary exercise test performed on a treadmill. A secondary aim was to explore whether measurement error varies across 5 predefined intensity zones (50%‐60%, 60%‐70%, 70%‐80%, 80%‐90%, and 90%‐100% of the maximum HR determined individually for each participant).
Methods
Overall, 55 healthy adults (30 men, 25 women; mean age 30.3, SD 8.2 years) completed a symptom-limited incremental treadmill protocol to volitional exhaustion. Simultaneous HR recordings were obtained from the GW6 (left arm) and a Polar H10 chest strap monitor, which served as the reference standards. For each intensity zone, the following agreement indices were computed: intraclass correlation coefficient (ICC), median absolute error, median absolute percentage error, and root mean squared error. Bland-Altman analysis was performed to quantify the mean bias and 95% limits of agreement between the GW6 and the Polar H10. Statistical significance was set at P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Health Data Lab, Sidia Institute of Science and Technology, Av. Autaz Mirim, 2211 - Distrito Industrial I, Manaus, 69075-155, Brazil, 55 (92) 3212-3444.
Zitieren
APA
Inoue, A., Soares, J. P. F., Antunes-Santos, F., Ferreira, A., Gonçalves, A., Alcântara, J. A., & Dos Santos, M. R. (2026). Heart Rate Estimation Using the Galaxy Watch During Maximal Cardiopulmonary Exercise Testing: Cross-Sectional Validation Study. JMIR Cardio, 10, e81917-e81917. https://doi.org/10.2196/81917
Vancouver
Inoue A, Soares JPF, Antunes-Santos F, Ferreira A, Gonçalves A, Alcântara JA, et al. Heart Rate Estimation Using the Galaxy Watch During Maximal Cardiopulmonary Exercise Testing: Cross-Sectional Validation Study. JMIR Cardio. 2026;10:e81917-e81917. doi:10.2196/81917
BibTeX
@article{Inoue2026,
author = {Inoue, Allan and Soares, João Paulo Ferreira and Antunes-Santos, Felipe and Ferreira, Alexandre and Gonçalves, Alberto and Alcântara, João Arthur and Dos Santos, Marcelo Rodrigues},
title = {Heart Rate Estimation Using the Galaxy Watch During Maximal Cardiopulmonary Exercise Testing: Cross-Sectional Validation Study},
journal = {JMIR Cardio},
year = {2026},
volume = {10},
pages = {e81917--e81917},
doi = {10.2196/81917}
}
EUVIMED