Parameters Influencing the Accuracy of a Wrist Photoplethysmography Heart-Rate Monitor (Polar Unite) During Exercise
International Journal of Sports Physiology and Performance · 2023
DOI: 10.1123/ijspp.2022-0288 ↗Kurz gefasst
Klinische Studie in der Zeitschrift International Journal of Sports Physiology and Performance, 2023. Verfasst von Quentin Bretonneau und 3 weiteren Autor:innen. Beteiligte Einrichtung: Laboratoire MOVE (UR 20296), Faculté des sciences du sport, Université de Poitiers, Poitiers,France. Hauptthemen laut MEDLINE: Photoplethysmography, Wrist. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 6 Literaturangaben.
Abstract (Originalsprache)
Purpose
The accuracy of heart rate measured with a wrist photoplethysmography monitor can be influenced by the tightening of the wristband, movement of arms, or kinetics of the signal (eg, steady-state exercise vs on- and off-transients). To test these hypotheses, photoplethysmographic and electrocardiographic (ECG) signals were compared.
Methods
Thirty participants (50% female) randomly performed two 13′ sequences (3′ rest, 5′ submaximal-intensity exercise, and 5′ passive recovery) on a motorized treadmill and a bicycle ergometer. Heart rate was measured concomitantly with a 10-lead ECG, a chest-strap monitor, and 2 wrist photoplethysmography monitors (Polar Unite) with different tightening (free vs imposed at the maximum tolerable).
Results
The level of association ( r ) and coefficient of variation (CV; ie, the error of measurement) of the Polar Unite versus the 10-lead ECG is affected by the tightness of the wristband (normal vs high; r = .83 and .96, CV = 16.1 and 8.1% for the treadmill, respectively; r = .71 and .97, CV = 20.3% and 6.2% for the bicycle, respectively) by the phase of the signal (transition vs steady state; r = .90 and .97, CV = 9.0% and 7.6% for the treadmill, respectively; r = .93 and .99, CV = 7.5% and 3.1% for the bicycle, respectively) and movement of arms (treadmill vs bicycle; r = .90 and .93, CV = 9.0% and 7.5% during the transition phase, respectively; r = .97 and .99, CV = 7.6% and 3.1% during the steady-state phase, respectively).
Conclusion
The accuracy of heart rate measured with a wrist photoplethysmography monitor is affected by the tightness of the wristband and the phase of the signal. A high tightening is required when high accuracy is expected.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratoire MOVE (UR 20296), Faculté des sciences du sport, Université de Poitiers, Poitiers,France.
ORCID-Profile: Quentin Bretonneau, Laurent Bosquet
Zitieren
APA
Bretonneau, Q., Peruque-Gayou, E., Wolfs, E., & Bosquet, L. (2023). Parameters Influencing the Accuracy of a Wrist Photoplethysmography Heart-Rate Monitor (Polar Unite) During Exercise. International Journal of Sports Physiology and Performance. https://doi.org/10.1123/ijspp.2022-0288
Vancouver
Bretonneau Q, Peruque-Gayou E, Wolfs E, Bosquet L. Parameters Influencing the Accuracy of a Wrist Photoplethysmography Heart-Rate Monitor (Polar Unite) During Exercise. International Journal of Sports Physiology and Performance. 2023. doi:10.1123/ijspp.2022-0288
BibTeX
@article{Bretonneau2023,
author = {Bretonneau, Quentin and Peruque-Gayou, Etienne and Wolfs, Etienne and Bosquet, Laurent},
title = {Parameters Influencing the Accuracy of a Wrist Photoplethysmography Heart-Rate Monitor (Polar Unite) During Exercise},
journal = {International Journal of Sports Physiology and Performance},
year = {2023},
doi = {10.1123/ijspp.2022-0288}
}
EUVIMED