EUVIMEDEUVIMED entdecken
VergleichsstudieEvaluationsstudieValidierungsstudieOpen Access

Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study

Femke Wouters iD, Henri Gruwez iD, Christophe Smeets iD, Anessa Pijalovic iD, Wouter Wilms iD, Julie Vranken iD, Zoë Pieters iD, Hugo Van Herendael iD, Dieter Nuyens iD, Maximo Rivero-Ayerza iD, Pieter Vandervoort iD, Peter Haemers iD und 1 weitere

JMIR Formative Research · 09.01.2025 · 9 · e65139-e65139

DOI: 10.2196/65139 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift JMIR Formative Research, online erschienen am 09.01.2025. Verfasst von Femke Wouters und 12 weiteren Autor:innen. Beteiligte Einrichtung: Limburg Clinical Research Center/Mobile Health Unit, Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Arrhythmias, Cardiac, Atrial Fibrillation, Electrocardiography. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Consumer-oriented wearable devices (CWDs) such as smartphones and smartwatches have gained prominence for their ability to detect atrial fibrillation (AF) through proprietary algorithms using electrocardiography or photoplethysmography (PPG)–based digital recordings. Despite numerous individual validation studies, a direct comparison of interdevice performance is lacking.

Objective

This study aimed to evaluate and compare the ability of CWDs to distinguish between sinus rhythm and AF.

Methods

Patients exhibiting sinus rhythm or AF were enrolled through a cardiology outpatient clinic. The participants were instructed to perform heart rhythm measurements using a handheld 6-lead electrocardiogram (ECG) device (KardiaMobile 6L), a smartwatch-derived single-lead ECG (Apple Watch), and two PPG-based smartphone apps (FibriCheck and Preventicus) in a random sequence, with simultaneous 12-lead reference ECG as the gold standard.

Results

A total of 122 participants were included in the study: median age 69 (IQR 61-77) years, 63.9% (n=78) men, 25% (n=30) with AF, 9.8% (n=12) without prior smartphone experience, and 73% (n=89) without experience in using a smartwatch. The sensitivity to detect AF was 100% for all devices. The specificity to detect sinus rhythm was 96.4% (95% CI 89.5%-98.8%) for KardiaMobile 6L, 97.8% (95% CI 91.6%‐99.5%) for Apple Watch, 98.9% (95% CI 92.5%‐99.8%) for FibriCheck, and 97.8% (95% CI 91.5%‐99.4%) for Preventicus ( P =.50). Insufficient quality measurements were observed in 10.7% (95% CI 6.3%-17.5%) of cases for both KardiaMobile 6L and Apple Watch, 7.4% (95% CI 3.9%‐13.6%) for FibriCheck, and 14.8% (95% CI 9.5%‐22.2%) for Preventicus ( P =.21). Participants preferred Apple Watch over the other devices to monitor their heart rhythm.

Conclusions

In this study population, the discrimination between sinus rhythm and AF using CWDs based on ECG or PPG was highly accurate, with no significant variations in performance across the examined devices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesArrhythmias, CardiacAtrial FibrillationElectrocardiographyPhotoplethysmographyHumansMaleFemaleMiddle AgedAgedMobile ApplicationsRandom AllocationSensitivity and SpecificityPatient PreferenceReproducibility of ResultsHeart Conduction System

Stichwörter der Autor:innen

AFECGatrial fibrillationconsumer wearable devicedetectionelectrocardiogramelectrocardiographymHealthmobile healthphotoplethysmographysmartphonewearable devices

Autor:innen und Institutionen

  • Limburg Clinical Research Center/Mobile Health Unit, Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium.
  • Department Future Health, Ziekenhuis Oost-Limburg, Genk, Belgium.
  • Department of Cardiology, Ziekenhuis Oost-Limburg, Genk, Belgium.
  • Department of Cardiovascular Sciences, KU Leuven, Leuven, Belgium.
  • Data Science Institute, Hasselt University, Hasselt, Belgium.

ORCID-Profile: Femke Wouters, Henri Gruwez, Christophe Smeets, Anessa Pijalovic, Wouter Wilms, Julie Vranken, Zoë Pieters, Hugo Van Herendael, Dieter Nuyens, Maximo Rivero-Ayerza, Pieter Vandervoort, Peter Haemers, Laurent Pison

Zitieren

APA

Wouters, F., Gruwez, H., Smeets, C., Pijalovic, A., Wilms, W., Vranken, J., Pieters, Z., Van Herendael, H., Nuyens, D., Rivero-Ayerza, M., Vandervoort, P., Haemers, P., & Pison, L. (2025). Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study. JMIR Formative Research, 9, e65139-e65139. https://doi.org/10.2196/65139

Vancouver

Wouters F, Gruwez H, Smeets C, Pijalovic A, Wilms W, Vranken J, et al. Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study. JMIR Formative Research. 2025;9:e65139-e65139. doi:10.2196/65139

BibTeX
@article{Wouters2025,
  author = {Wouters, Femke and Gruwez, Henri and Smeets, Christophe and Pijalovic, Anessa and Wilms, Wouter and Vranken, Julie and Pieters, Zoë and Van Herendael, Hugo and Nuyens, Dieter and Rivero-Ayerza, Maximo and Vandervoort, Pieter and Haemers, Peter and Pison, Laurent},
  title = {Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study},
  journal = {JMIR Formative Research},
  year = {2025},
  volume = {9},
  pages = {e65139--e65139},
  doi = {10.2196/65139}
}
Kostenlos starten →