Integrated Support Surface With Multimodal Sensing and Low-Flow Oxygen Delivery: Design and Bench Characterization (Preprint)

Taryudi Taryudi, Linlin Lindayani, Yen-Chin Chen, Taufiqqurrachman Taufiqqurrachman, Astri Mutiar, Irma Darmawati

JMIR Biomedical Engineering · 2026-08-27

DOI: 10.2196/110611 ↗

At a glance

Journal article in JMIR Biomedical Engineering, published online on 2026-08-27. Written by Taryudi Taryudi and 5 co-authors. 19 references.

AbstractDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Pressure injury risk reflects sustained tissue deformation together with impaired tissue tolerance and adverse skin-support microclimate. Continuous monitoring can characterize interface conditions over time. We developed an investigational smart bedding surface that combines distributed pressure, temperature, and relative-humidity sensing with controlled low-flow oxygen delivery. This study evaluated short-duration technical performance and formative simulated-use feasibility; it did not test clinical safety or pressure injury prevention effectiveness.

Results

At nominal reference loads of 40, 80, and 120 mmHg, mean pressure readings were 41.2 ± 1.5, 78.9 ± 2.1, and 118.4 ± 2.8 mmHg, respectively. The absolute difference between each measured mean and its reference value was 1.1–1.6 mmHg, and coefficients of variation were 2.4%-3.6%. The source summary reported mean temperature and relative-humidity errors of 0.3°C and 2.1% RH, respectively. During 6-, 12-, and 24-h operation, mean oxygen flow was 0.98–1.01 L/min against a 1.0-L/min target; surface temperature increased by 0.3–0.5°C and relative humidity by 1.8%-2.4%. Reported 24-h changes were + 1.9% for pressure, + 2.3% for relative humidity, and + 0.20°C for temperature. In formative simulated-use tasks, mean setup, daily-check, and cleaning times were 6.5, 2.1, and 4.3 min, respectively; cable routing was the main observed operational issue.

Conclusions

The available bench data support the engineering feasibility of integrating multimodal interface sensing and low-flow oxygen delivery in one bedding platform. They do not establish clinical efficacy, comprehensive safety, metrological agreement across use conditions, or readiness for patient use. Replicated testing against traceable reference instruments, verification of oxygen distribution and oxygen-related hazards, formal usability engineering, durability testing, and staged clinical evaluation are required.

Cite

APA

Taryudi, T., Lindayani, L., Chen, Y., Taufiqqurrachman, T., Mutiar, A., & Darmawati, I. (2026). Integrated Support Surface With Multimodal Sensing and Low-Flow Oxygen Delivery: Design and Bench Characterization (Preprint). JMIR Biomedical Engineering. https://doi.org/10.2196/110611

Vancouver

Taryudi T, Lindayani L, Chen Y, Taufiqqurrachman T, Mutiar A, Darmawati I. Integrated Support Surface With Multimodal Sensing and Low-Flow Oxygen Delivery: Design and Bench Characterization (Preprint). JMIR Biomedical Engineering. 2026. doi:10.2196/110611

BibTeX
@article{Taryudi2026,
  author = {Taryudi, Taryudi and Lindayani, Linlin and Chen, Yen-Chin and Taufiqqurrachman, Taufiqqurrachman and Mutiar, Astri and Darmawati, Irma},
  title = {Integrated Support Surface With Multimodal Sensing and Low-Flow Oxygen Delivery: Design and Bench Characterization (Preprint)},
  journal = {JMIR Biomedical Engineering},
  year = {2026},
  doi = {10.2196/110611}
}
Start for free →