Delivery of high flow oxygen through nasal vs. tracheal cannulas: A bench study
Frontiers in Medicine · 12.01.2023 · 9
DOI: 10.3389/fmed.2022.1068428 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 12.01.2023. Verfasst von Martin Cour und 4 weiteren Autor:innen. Beteiligte Einrichtung: Service de Médecine Intensive-Réanimation, Hospices Civils de Lyon, Hôpital Edouard Herriot, Lyon, France. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The use of high flow oxygen therapy (HFOT) has significantly escalated during the COVID-19 pandemic. HFOT can be delivered through both dedicated devices and ICU ventilators. HFOT can be administered to a patient via a nasal cannula (NC). In intubated patients, a tracheal cannula (TC) is used instead. In this study, we aim to compare the work of breathing (WOB) using a TC or NC and to explore whether differences exist among HFOT devices.
Methods
Seven HFOT devices (three dedicated and four ICU ventilators) were connected to a manikin head (Laerdal Medical) through a NC (Optiflow 3S, large size, Fisher and Paykel Healthcare) or a TC (OPT 970 Optiflow+, Fisher and Paykel Healthcare). Each device was also attached to a manikin head that was connected to a lung simulator (ASL5000, Ingmar Medical), set at 40 ml/cmH 2 O compliance, 10 cmH 2 O/L/s resistance, and sinusoidal inspiratory effort (muscular pressure 10 cmH 2 O, rate 30 breaths/min). HFOT was delivered at 40 L/min and at 21% inspired oxygen fraction. The total WOB per breath and its resistive and elastic components were automatically analyzed breath by breath over the last 20 breaths by using Campbell's diagram.
Results
The WOB and its resistive and elastic components were significantly lower with the TC than with the NC for every device, and systematically lower with the reference device than with others. These differences were, however, very small and may be not clinically relevant.
Conclusion
The WOB is lower with the TC than with the NC and with the reference device, compared with the most recent devices.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Service de Médecine Intensive-Réanimation, Hospices Civils de Lyon, Hôpital Edouard Herriot, Lyon, France.
- Faculté de Médecine Lyon-Est, Université de Lyon, Université Claude Bernard Lyon 1, Lyon, France.
- Institut Mondor de Recherches Biomédicales, INSERM 955 CNRS 7000, Créteil, France.
Zitieren
APA
Cour, M., Guérin, C., Degivry, F., Argaud, L., & Louis, B. (2023). Delivery of high flow oxygen through nasal vs. tracheal cannulas: A bench study. Frontiers in Medicine, 9. https://doi.org/10.3389/fmed.2022.1068428
Vancouver
Cour M, Guérin C, Degivry F, Argaud L, Louis B. Delivery of high flow oxygen through nasal vs. tracheal cannulas: A bench study. Frontiers in Medicine. 2023;9. doi:10.3389/fmed.2022.1068428
BibTeX
@article{Cour2023,
author = {Cour, Martin and Guérin, Claude and Degivry, Florian and Argaud, Laurent and Louis, Bruno},
title = {Delivery of high flow oxygen through nasal vs. tracheal cannulas: A bench study},
journal = {Frontiers in Medicine},
year = {2023},
volume = {9},
doi = {10.3389/fmed.2022.1068428}
}
EUVIMED