Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation

Toru Ishikawa, Iori Hasegawa, Hiroshi Hirosawa, Tsubasa Honmou, Nobuyuki Sakai, Takanori Igarashi, Shun Yamazaki, Takamasa Kobayashi, Toshifumi Sato, Akito Iwanaga, Tomoe Sano, Junji Yokoyama und 1 weitere

In Vivo · 2023

DOI: 10.21873/invivo.13179 ↗

Kurz gefasst

Fachartikel in der Zeitschrift In Vivo, 2023. Verfasst von Toru Ishikawa und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology, Saiseikai Niigata Hospital, Niigata, Japan; toruishi@ngt.saiseikai.or.jp. Hauptthemen laut MEDLINE: Catheter Ablation, Radiofrequency Ablation. Die Arbeit wurde bisher 1-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND/AIM: Ablating a spherical area during hepatocellular carcinoma ablation therapy is a very important issue. We aimed to determine the ablation area of bovine liver using various radiofrequency ablation (RFA) protocols. MATERIALS AND METHODS: Bovine liver (1-2 kg) was placed in an aluminum tray, which was punctured with STARmed VIVA 2.0 17-gauge (G) and 15-G electrodes using a current-carrying tip. Under the step-up or linear method, with an ablation time up to one break and RFA output stop, the size of the color change area (representing the thermally coagulated area) of the bovine liver was measured along the vertical and horizontal axes, and the ablated volume and total heat generated were calculated. RESULTS: 5-W per minute increases protocol resulted in greater horizontal and vertical diameters of the ablated area than 10-W per minute increases protocol under the step-up method. For 5-W and 10-W per minute increases under the step-up method, the aspect ratio was 0.81 and 0.67 with a 17-G electrode, and 0.73 and 0.69 with a 15-G electrode, respectively. For 5-W and 10-W increases under the linear method, the aspect ratio was 0.89 and 0.82, respectively. Sufficient ablation was obtained, with vertical and horizontal diameters of 50 mm and 43.50 mm, respectively. Although the ablation time was long, the watt output value at the break and average watt value were low. CONCLUSION: Gradual increase in output (5 W) using the step-up method yielded a more spherical ablation area, and longer ablation time in the linear method with a 15-G electrode could result in a more spherical ablation area in real clinical practice in humans. Future studies should examine concerns regarding long ablation times.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheter AblationRadiofrequency AblationHumansAnimalsCattleLiverTime FactorsElectrodes

Stichwörter der Autor:innen

15 gaugeAblation arealinear moderadiofrequency ablation

Autor:innen und Institutionen

  • Department of Gastroenterology, Saiseikai Niigata Hospital, Niigata, Japan; toruishi@ngt.saiseikai.or.jp.
  • Department of Clinical Engineering, Saiseikai Niigata Hospital, Niigata, Japan.
  • Department of Gastroenterology, Saiseikai Niigata Hospital, Niigata, Japan.

Zitieren

APA

Ishikawa, T., Hasegawa, I., Hirosawa, H., Honmou, T., Sakai, N., Igarashi, T., Yamazaki, S., Kobayashi, T., Sato, T., Iwanaga, A., Sano, T., Yokoyama, J., & Honma, T. (2023). Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation. In Vivo. https://doi.org/10.21873/invivo.13179

Vancouver

Ishikawa T, Hasegawa I, Hirosawa H, Honmou T, Sakai N, Igarashi T, et al. Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation. In Vivo. 2023. doi:10.21873/invivo.13179

BibTeX
@article{Ishikawa2023,
  author = {Ishikawa, Toru and Hasegawa, Iori and Hirosawa, Hiroshi and Honmou, Tsubasa and Sakai, Nobuyuki and Igarashi, Takanori and Yamazaki, Shun and Kobayashi, Takamasa and Sato, Toshifumi and Iwanaga, Akito and Sano, Tomoe and Yokoyama, Junji and Honma, Terasu},
  title = {Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation},
  journal = {In Vivo},
  year = {2023},
  doi = {10.21873/invivo.13179}
}
Kostenlos starten →