EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range

Jinsu An, Dong-Sung Won, Yubeen Park, Jung-Hoon Park, Ki-Hyeon Park, Ji-Ho Lee, Hyung-Sik Kim iD

PLOS ONE · 2024

DOI: 10.1371/journal.pone.0308691 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Jinsu An und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungche…. Hauptthemen laut MEDLINE: Radio Waves. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: by the Korea Medical Device Development Fund grant funded by the Korea government.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study reports the effects of changes in the waveform and frequency of radio frequency (RF) energy on the tissue ablation range. We developed a 70-watt RFA generator that provides sine and square waves and allows frequency control between 10 Hz and 500 kHz. The changes in the ablation range according to the waveform and frequency were observed using the developed generator. In the waveform variation test, the distance between the electrodes and the electrode type were changed for both waveforms with the frequency set to 500 kHz. In the frequency variation test, the waveform and electrode type were changed with the frequency set to 10, 100, and 500 kHz, while the distance between the electrodes was set to 20 mm. A fixed 45 voltage was applied using the bipolar method. RF energy was applied for 90 s in vitro. The temperature was regulated to not exceed 70°C. The ablation range was calculated using ImageJ software. The analysis results showed that the ablation range was larger with the square wave than with the sine wave and at 10 kHz than at 500 kHz. The developed generator can advance research on ablation area and depth in RF ablation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radio WavesElectrodesCatheter AblationRadiofrequency AblationAnimalsTemperature

Autor:innen und Institutionen

  • Department of Biomedical Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea.
  • Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Songpa-gu, Seoul, Republic of Korea.
  • Department of Gastroenterology, Asan Medical Center, University of Ulsan College of Medicine, Songpa-gu, Seoul, Republic of Korea.
  • Department of Convergence Medicine, Asan Medical Center, University of Ulsan College of Medicine, Songpa-gu, Seoul, Republic of Korea.
  • Department of Mechatronics Engineering, School of ICT Convergence Engineering, College of Science & Technology, Konkuk University, Chungju-si, Chungcheongbuk-do, Republic of Korea.

ORCID-Profile: Hyung-Sik Kim

Förderung

  • by the Korea Medical Device Development Fund grant funded by the Korea government

Zitieren

APA

An, J., Won, D. S., Park, Y., Park, J. H., Park, K. H., Lee, J. H., & Kim, H. S. (2024). Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range. PLOS ONE. https://doi.org/10.1371/journal.pone.0308691

Vancouver

An J, Won DS, Park Y, Park JH, Park KH, Lee JH, et al. Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range. PLOS ONE. 2024. doi:10.1371/journal.pone.0308691

BibTeX
@article{An2024,
  author = {An, Jinsu and Won, Dong-Sung and Park, Yubeen and Park, Jung-Hoon and Park, Ki-Hyeon and Lee, Ji-Ho and Kim, Hyung-Sik},
  title = {Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range},
  journal = {PLOS ONE},
  year = {2024},
  doi = {10.1371/journal.pone.0308691}
}
Kostenlos starten →