In vitro study of optimal removable partial denture clasp design made from novel high-performance polyetherketoneketone

Pei-Wen Peng, May-Show Chen, Tzu-Yu Peng, Peng-Chien Huang, Hiroki Nikawa, Wei-Fang Lee

Journal of Prosthodontic Research · 2024

DOI: 10.2186/jpr.jpr_d_23_00080 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Prosthodontic Research, 2024. Verfasst von Pei-Wen Peng und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Dental Technology, Taipei Medical University, Taipei, Taiwan. Hauptthemen laut MEDLINE: Denture, Partial, Removable, Computer-Aided Design, Dental Clasps, Denture Design. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This study aimed to investigate the retentive force and deformation of double Akers' polyetherketoneketone (PEKK) clasps on removable partial dentures (RPDs) with varying designs and undercut depths. METHODS: Thirty double Akers' PEKK clasps with two different widths and heights (Groups I and II) were fabricated using computer-aided design and computer-aided manufacturing (CAD/CAM). Each design was further subdivided (n = 5) into three undercut depths (0.25, 0.50, and 0.75 mm). The retentive force of the clasps was measured after 10 years of clinical use (15,000 insertion/removal cycles), and the deformation of the clasp tips was analyzed before and after cycling. RESULTS: Clasps with 0.50-mm and 0.75-mm undercut depths exhibited greater initial retentive forces (3.15-3.51 N) compared to those in the 0.25-mm undercut group (2.40-2.80 N). Group I maintained consistent retentive forces over the cycles (P = 0.345), whereas Group II showed declining forces after the initial use (P < 0.003). In both groups, the 0.50-mm undercut exhibited a greater retentive force than the 0.25-mm (P < 0.001 and P < 0.004, respectively), with no significant differences between the 0.50-mm and 0.75-mm undercut depths. Despite a lower initial retentive force, the 0.25-mm undercut showed less deformation and clasp tip wear. CONCLUSIONS: The PEKK clasps did not exhibit significantly reduced retentive forces or permanent deformations after 15,000 fatigue cycles. These results suggest that the PEKK polymer displays superior mechanical properties as an esthetic clasp material, and clasps with 0.50-mm and 0.75-mm undercut depths are recommended for long-term clinical use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Denture, Partial, RemovableComputer-Aided DesignDental ClaspsDenture DesignKetonesPolymersDenture RetentionBenzophenonesMaterials TestingIn Vitro TechniquesDental Stress Analysis

Stichwörter der Autor:innen

ClaspPEKKRetentive forceRremovable partial denture

Substanzen

KetonesPolymerspolyetherketoneketoneBenzophenones

Autor:innen und Institutionen

  • School of Dental Technology, Taipei Medical University, Taipei, Taiwan.
  • School of Dentistry, Taipei Medical University, Taipei, Taiwan.
  • Division of Prosthodontics, Department of Dentistry, Taipei Medical University Hospital, Taipei, Taiwan.
  • Department of Oral Biology & Engineering, Graduate School of Biomedical and Health Sciences, Hiroshima University, Hiroshima City, Hiroshima, Japan.

Zitieren

APA

Peng, P. W., Chen, M. S., Peng, T. Y., Huang, P. C., Nikawa, H., & Lee, W. F. (2024). In vitro study of optimal removable partial denture clasp design made from novel high-performance polyetherketoneketone. Journal of Prosthodontic Research. https://doi.org/10.2186/jpr.jpr_d_23_00080

Vancouver

Peng PW, Chen MS, Peng TY, Huang PC, Nikawa H, Lee WF. In vitro study of optimal removable partial denture clasp design made from novel high-performance polyetherketoneketone. Journal of Prosthodontic Research. 2024. doi:10.2186/jpr.jpr_d_23_00080

BibTeX
@article{Peng2024,
  author = {Peng, Pei-Wen and Chen, May-Show and Peng, Tzu-Yu and Huang, Peng-Chien and Nikawa, Hiroki and Lee, Wei-Fang},
  title = {In vitro study of optimal removable partial denture clasp design made from novel high-performance polyetherketoneketone},
  journal = {Journal of Prosthodontic Research},
  year = {2024},
  doi = {10.2186/jpr.jpr_d_23_00080}
}
Kostenlos starten →