Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol
Vascular Health and Risk Management · 2025
DOI: 10.2147/vhrm.s513013 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Vascular Health and Risk Management, 2025. Verfasst von Zoya Hakobyan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Center of Preventive Cardiology, Hospital N2 CJSC, Yerevan, Armenia. Hauptthemen laut MEDLINE: Blood Pressure Determination, Blood Pressure, Wrist, Oscillometry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0).
Abstract (Originalsprache)
Background
Scientific societies universally recommend evaluating the accuracy of electronic devices designed for blood pressure (BP) measurement using established validation protocols.
Objective
This study aimed to assess the accuracy of the Combei BP880W wrist device for BP measurement in the general population, according to the ISO 81060-2:2018/AMD 1:2020 Universal Standard.
Methods
The Combei BP880W is an oscillometric device designed to measure BP at the wrist. This study adhered to the ISO 81060-2:2018/AMD 1:2020 protocol and employed the same-arm sequential BP measurement method. A total of 85 participants, meeting protocol-specified age, gender, BP, and cuff distribution criteria, were included. The accuracy analysis utilized Criterion 1 (differences and standard deviations between reference and test device measurements) and Criterion 2 (intra-individual standard deviation of BP differences).
Results
Eighty-five participants were included. Mean BP differences between the simultaneous observer measurements were -0.2 ± 1.9 mmHg for systolic BP (SBP) and 0.1 ± 1.9 mmHg for diastolic BP (DBP). For Criterion 1, the mean difference ± standard deviation (SD) between the reference and test device measurements were -2.7 ± 5.9 mmHg (SBP) and -2.0 ± 3.9 mmHg (DBP), meeting the required threshold (≤ 5 ± 8 mmHg). For Criterion 2, intra-individual SDs were 4.6 mmHg (SBP) and 3.4 mmHg (DBP), both below the respective limits (≤ 6.39 mmHg for SBP and ≤ 6.65 mmHg for DBP).
Conclusion
The Combei BP880W wrist device meets the accuracy requirements of the ISO 81060-2:2018/AMD 1:2020 protocol, supporting its use for home BP monitoring in the general population.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center of Preventive Cardiology, Hospital N2 CJSC, Yerevan, Armenia.
- The Institute of Cardiology, Yerevan, Armenia.
- Diagnosis and Therapeutic Center, Hôtel Dieu Hospital, Paris, France.
- Foundation-Medical Research Institutes (F-MRI) ®, Geneva, Switzerland.
ORCID-Profile: Zoya Hakobyan, Roland Asmar
Zitieren
APA
Hakobyan, Z., Zelveian, P., Gharibyan, H., Topouchian, J., & Asmar, R. (2025). Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol. Vascular Health and Risk Management. https://doi.org/10.2147/vhrm.s513013
Vancouver
Hakobyan Z, Zelveian P, Gharibyan H, Topouchian J, Asmar R. Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol. Vascular Health and Risk Management. 2025. doi:10.2147/vhrm.s513013
BibTeX
@article{Hakobyan2025,
author = {Hakobyan, Zoya and Zelveian, Parounak and Gharibyan, Heghine and Topouchian, Jirar and Asmar, Roland},
title = {Accuracy of the Combei BP880W Wrist Device for Self-Blood Pressure Measurements in General Population According to the International Organization for Standardization Universal Standard (ISO 81060-2:2018/AMD 1:2020) Protocol},
journal = {Vascular Health and Risk Management},
year = {2025},
doi = {10.2147/vhrm.s513013}
}
EUVIMED