Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract

Shan Ma, Cheng Li, Jing Sun, Jun Yang, Kai Wen, Xi-Teng Chen, Fang-Yu Zhao, Rong-Yu Gao, Fang Tian

International Journal of Ophthalmology · 18.08.2024 · 17 (8) · 1437-1446

DOI: 10.18240/ijo.2024.08.08 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Ophthalmology, erschienen am 18.08.2024. Verfasst von Shan Ma und 8 weiteren Autor:innen. Beteiligte Einrichtung: Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and S…. Die Arbeit wurde bisher 8-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

AIM: To assess and compare the variations and agreements across different ocular biometric parameters using swept-source optical coherence tomography (SS-OCT) and Scheimpflug tomography in patients diagnosed with cataract. METHODS: This prospective case series was conducted at Tianjin Medical University Eye Hospital. In total, 212 eyes from 212 patients scheduled for phacoemulsification were included. Eyes were evaluated preoperatively using two SS-OCT devices (IOLMaster700 and CASIA2) and Scheimpflug tomography (Pentacam). Central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), white-to-white distance (WTW), flat simulated keratometry (Kf), steep simulated keratometry (Ks), mean keratometry (Km), and total corneal keratometry (TKm) were measured. Intraclass correlation coefficient (ICC), 95% confidence intervals (CI) and limits of agreement (LoA) widths were conducted to assess differences and correlations between devices. RESULTS: All parameters, except for Ks, were significantly different. Pairwise comparison revealed no significant differences between keratometry obtained by IOLMaster 700 and Pentacam. LoA widths of all paired comparisons for Ks were >0.80 D. Except for WTW between IOLMaster 700 and CASIA2 and between CASIA2 and Pentacam, other Pearson’s coefficients between devices showed a strong correlation (all r>0.95). The ICC of WTW (ICC=0.438, 95%CI 0.167-0.625) showed poor reliability. The reliability of CCT, ACD, and AQD was excellent (all ICC>0.95), whereas that of TKm was good (ICC=0.827, 95%CI 0.221-0.939). A significant linear correlation was also observed among devices. CONCLUSION: The ocular parameters derived from the use of IOLMaster700, CASIA2, and Pentacam exhibit significant discrepancies; as such, measurements from these devices should not be deemed as interchangeable.

Schlagwörter

Stichwörter der Autor:innen

cataractocular parametersswept-source optical coherence tomography

Autor:innen und Institutionen

  • Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 3…
  • Zhengda Guangming Eye Group, Weifang 261000, Shandong Province, China.
  • Binzhou Polytechnic, Binzhou 256603, Shandong Province, China.

Zitieren

APA

Ma, S., Li, C., Sun, J., Yang, J., Wen, K., Chen, X. T., Zhao, F. Y., Gao, R. Y., & Tian, F. (2024). Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract. International Journal of Ophthalmology, 17(8), 1437-1446. https://doi.org/10.18240/ijo.2024.08.08

Vancouver

Ma S, Li C, Sun J, Yang J, Wen K, Chen XT, et al. Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract. International Journal of Ophthalmology. 2024;17(8):1437-1446. doi:10.18240/ijo.2024.08.08

BibTeX
@article{Ma2024,
  author = {Ma, Shan and Li, Cheng and Sun, Jing and Yang, Jun and Wen, Kai and Chen, Xi-Teng and Zhao, Fang-Yu and Gao, Rong-Yu and Tian, Fang},
  title = {Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract},
  journal = {International Journal of Ophthalmology},
  year = {2024},
  volume = {17},
  number = {8},
  pages = {1437--1446},
  doi = {10.18240/ijo.2024.08.08}
}
Kostenlos starten →