Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract

Shan Ma, Rongyu Gao, Jing Sun, Jun Yang, Kai Wen, Xiteng Chen, Fangyu Zhao, Xinyan Xu, Fang Tian

Graefe's Archive for Clinical and Experimental Ophthalmology · 2024-05 · 262 (5) · 1567-1578

DOI: 10.1007/s00417-023-06348-y ↗

Kurz gefasst

Fachartikel in der Zeitschrift Graefe's Archive for Clinical and Experimental Ophthalmology, online erschienen am 27.12.2023. Verfasst von Shan Ma und 8 weiteren Autor:innen. Beteiligte Einrichtung: Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and S…. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Cataract. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tianjin Key Medical Discipline (Specialty) Construction Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess the differences and similarities in the corneal curvature obtained by two swept-source optical coherence tomography (SS-OCT) devices, Scheimpflug imaging system and one ray tracing aberrometer in patients with cataracts. Moreover, this study aimed to compare the differences in posterior corneal (PK), total corneal (TK) and true net power (TNP) measurements among the IOLMaster 700, CASIA2, and Pentacam.

Methods

A total of 200 eyes of 200 patients (116 female, 58%) were enrolled in this study, with a mean age of 65.9 ± 9.5 years. The flattest (Kf), steepest (Ks), and mean cornal powers (Km), J 0 , and J 45 were obtained using two SS-OCT-based biometric devices, one rotating camera system and one ray-tracing aberrometer. The PK, TK and TNP values were also measured using these devices. To evaluate the differences and similarities between the devicves, the Friedman test, Pearson correlation coefficient (r), intraclass coefficient correlation (ICC) and Bland‒Altman plots with 95% limits of agreement (LoA) were used, and boxplots and stacked histograms were generated to describe the distributions of the data.

Results

There were no significant differences between the IOLMaster 700 and Pentacam for any of the keratometry values. Additionally, there were no significant differences between the IOLMaster 700 and iTrace in evaluating J 0 and J 45 . Bland‒Altman plots revealed relatively wide LoA widths, almost larger than 1 diopter for the keratometry values and almost larger than 0.5 diopter for J 0 and J 45 values among the four devices. In terms of PK and TK values, significant differences and low ICCs were found among the three devices.

Conclusions

Although strong correlations and good agreement were found among the IOLMaster700, CASIA2, Pentacam and iTrace for Kf, Ks, Km and J 0 , J 45 , it seems that the measurements should not be used interchangeably because of the wide LoA widths and the presence of significant differences among the devices. Similarly, due to significant differences and low ICCs, the PK, TK and TNP values obtained by IOLMaster 700, CASIA2, and Pentacam should not be used interchangeably.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceCataractHumansFemaleMiddle AgedAgedProspective StudiesReproducibility of ResultsCorneaBiometryCorneal Topography

Stichwörter der Autor:innen

Astigmatic vector analysisCASIA2Corneal curvatureIOLMaster 700PentacamiTrace

Autor:innen und Institutionen

  • Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin,…
  • Weifang Eye Hospital, National Key Clinical Specialty, Zhengda Guangming Eye Group, Weifang, 261000, China.
  • Weifang Eye Institute, Weifang, 261000, China.
  • Weifang Eye Hospital, National Key Clinical Specialty, Zhengda Guangming Eye Group, Weifang, 261000, China. zczdgmykyy2353@163.com.
  • Weifang Eye Institute, Weifang, 261000, China. zczdgmykyy2353@163.com.

Förderung

  • Tianjin Key Medical Discipline (Specialty) Construction Project

Zitieren

APA

Ma, S., Gao, R., Sun, J., Yang, J., Wen, K., Chen, X., Zhao, F., Xu, X., & Tian, F. (2024). Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract. Graefe's Archive for Clinical and Experimental Ophthalmology, 262(5), 1567-1578. https://doi.org/10.1007/s00417-023-06348-y

Vancouver

Ma S, Gao R, Sun J, Yang J, Wen K, Chen X, et al. Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract. Graefe's Archive for Clinical and Experimental Ophthalmology. 2024;262(5):1567-1578. doi:10.1007/s00417-023-06348-y

BibTeX
@article{Ma2024,
  author = {Ma, Shan and Gao, Rongyu and Sun, Jing and Yang, Jun and Wen, Kai and Chen, Xiteng and Zhao, Fangyu and Xu, Xinyan and Tian, Fang},
  title = {Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract},
  journal = {Graefe's Archive for Clinical and Experimental Ophthalmology},
  year = {2024},
  volume = {262},
  number = {5},
  pages = {1567--1578},
  doi = {10.1007/s00417-023-06348-y}
}
Kostenlos starten →