Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography

Kamini Narendra Reddy, Ramadhan Ahmed iD, Matthew Zeng iD, Nakul S Shekhawat iD

Bioengineering · 21.08.2026 · 13 (8) · 944

DOI: 10.3390/bioengineering13080944 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 21.08.2026. Verfasst von Kamini Narendra Reddy und 3 weiteren Autor:innen. Beteiligte Einrichtung: Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, Raab Family Rising Professorship, Heidelberg Engineering (Germany).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare concordance of corneal tomographic measurements obtained using Anterion versus Pentacam HR.

Setting

Outpatient ophthalmology clinic.

Design

Retrospective cross-sectional study.

Methods

Eyes without corneal abnormalities underwent tomography using the Anterion Cornea App and Pentacam HR. Measurements included anterior keratometry (K1, K2, ΔK, average K, Kmax), posterior keratometry (K1, K2, ΔK, and average K), astigmatism power-vector components (J0, J45) for both surfaces, central corneal thickness (CCT), white-to-white (WTW), and thinnest pachymetric point (TP). We assessed agreement using Pearson correlation coefficients (ρ), Lin’s concordance correlation coefficient (CCC), and Bland–Altman analysis. Generalized estimating equation models were used to assess measurement differences between devices.

Results

A total of 64 eyes of 41 patients underwent corneal tomography with both devices. Anterion and Pentacam HR showed strong linear correlations for all measurements (ρ > 0.700), with concordance ranging from satisfactory to excellent (CCC 0.704–0.981). No statistically significant mean inter-device differences were found for anterior K1 (0.01 diopters [D], p = 0.775), K2 (−0.003 D, p = 0.951), average K (0.01 D, p = 0.801), ΔK (−0.02 D, p = 0.693), or the anterior astigmatism vectors J0 (0.002 D, p = 0.936) and J45 (0.03 D, p = 0.129). Mean posterior K1, K2, and average K measurements were flatter with Anterion than Pentacam HR, with mean inter-device differences of 0.13 D (p < 0.001), 0.11 D (p < 0.001), and 0.12 D (p < 0.001), respectively. The mean inter-device difference for posterior ΔK was −0.02 D (p = 0.018). Statistically significant mean inter-device differences were also found for posterior J0 (−0.01 D, p = 0.018) and J45 (0.01 D, p = 0.007). Anterion measured lower values than Pentacam HR for anterior Kmax, CCT, and TP, with mean inter-device differences of −0.24 D (p = 0.002), −7.66 μm (p < 0.001), and −8.58 μm (p < 0.001), respectively, and a higher mean WTW, with a mean inter-device difference of 0.07 mm (p < 0.001).

Conclusions

Anterion and Pentacam HR showed satisfactory-to-excellent measurement concordance, with negligible mean inter-device biases for most anterior keratometry measurements and anterior astigmatism vectors. However, individual-eye differences could approach 0.49–0.61 D for anterior keratometry and exceed 20 μm for pachymetry. These differences may become relevant in borderline cases and during longitudinal disease monitoring, warranting parameter-specific caution.

Schlagwörter

Stichwörter der Autor:innen

Scheimpflug tomographycorneapachymetryswept source optical coherence tomography

Autor:innen und Institutionen

  • Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA.
  • School of Medicine, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Department of Biology, Krieger School of Arts and Sciences, Johns Hopkins University, Baltimore, MD 21218, USA.

ORCID-Profile: Ramadhan Ahmed, Matthew Zeng, Nakul S Shekhawat

Förderung

  • NEI NIH HHS
  • Raab Family Rising Professorship
  • Heidelberg Engineering (Germany)
  • National Eye Institute

Zitieren

APA

Reddy, K. N., Ahmed, R., Zeng, M., & Shekhawat, N. S. (2026). Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography. Bioengineering, 13(8), 944. https://doi.org/10.3390/bioengineering13080944

Vancouver

Reddy KN, Ahmed R, Zeng M, Shekhawat NS. Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography. Bioengineering. 2026;13(8):944. doi:10.3390/bioengineering13080944

BibTeX
@article{Reddy2026,
  author = {Reddy, Kamini Narendra and Ahmed, Ramadhan and Zeng, Matthew and Shekhawat, Nakul S},
  title = {Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {8},
  pages = {944},
  doi = {10.3390/bioengineering13080944}
}
Kostenlos starten →