EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC-ND 4.0

Comparison of immune responses to zirconia, polyether ether ketone (PEEK), and stainless-steel in orthopaedic implants

Agung Riyanto Budi Santoso, Respati Suryanto Dradjat, Edi Mustamsir, Setyawan Purnomo Sakti, Panji Sananta, Hidayat Sujuti, Heri Suroto

Medicinski Glasnik · 2026

DOI: 10.17392/2037-23-01 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Medicinski Glasnik, 2026. Verfasst von Agung Riyanto Budi Santoso und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic and Traumatology, RSUD Prof. Dr. Saiful Anwar Malang, Malang, Indonesia. Hauptthemen laut MEDLINE: Zirconium, Stainless Steel, Polyethylene Glycols, Ketones. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

<p><strong>Aim </strong>Orthopedic implants must meet specific criteria, including mechanical strength, durability, and biocompatibility. This study compares the immune response of zirconia, polyether ether ketone (PEEK), and stainless-steel implants <em>in vivo</em>, focusing on lymphocyte and fibroblast infiltration as indicators of immune activation.</p> <p><strong>Methods</strong><strong> </strong>A total of 27 New Zealand white rabbits were used, with nine animals in each group. Implants of zirconia, PEEK, or stainless steel were surgically placed in the thigh and observed for 4 weeks. Histological analysis measured lymphocyte and fibroblast infiltration at the implant site using a microscope at 400x magnification. Statistical analysis included the Kruskal-Wallis test for group comparisons, followed by Mann-Whitney and Bonferroni correction for pairwise comparisons.</p> <p><strong>Results </strong>The Kruskal-Wallis test showed significant differences in lymphocyte (p=0.002) and fibroblast (p=0.003) counts among the groups. Zirconia exhibited significantly lower lymphocyte (median=0.5) and fibroblast (median=1.0) infiltration compared to stainless steel (lymphocytes: median=3.0, fibroblasts: median=2.0), and PEEK (lymphocytes: median=2.0, fibroblasts: median=3.0). Bonferroni correction confirmed zirconia showed the least immune activation (p&lt;0.0167).</p> <p><strong>Conclusion </strong>Zirconia offers superior biocompatibility with minimal immune response, making it an ideal material for orthopedic implants, particularly for patients with metal sensitivities. PEEK showed moderate immune activation but is helpful for non-load-bearing applications. Stainless Steel induced the highest immune response due to the release of metal ions and corrosion. Zirconia is the most biocompatible material tested, making it a promising choice for orthopedic implants.</p>

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumStainless SteelPolyethylene GlycolsKetonesProstheses and ImplantsBiocompatible MaterialsAnimalsRabbitsBenzophenonesPolyether CompoundsPolymersFibroblastsMaterials TestingLymphocytes

Stichwörter der Autor:innen

immunitypolyetheretherketoneprostheses and implantsstainless-steelzirconia

Substanzen

ZirconiumStainless Steelzirconium oxideBenzophenonespolyetheretherketonePolyethylene GlycolsKetonesPolyether CompoundsBiocompatible MaterialsPolymers

Autor:innen und Institutionen

  • Department of Orthopedic and Traumatology, RSUD Prof. Dr. Saiful Anwar Malang, Malang, Indonesia.
  • Department of Physics, University of Brawijaya, Malang, Indonesia.
  • Department of Biochemistry and Biomolecular, University of Brawijaya, Malang, Indonesia.
  • Department of Orthopedic and Traumatology, Rumah Sakit Dokter Soetomo, Surabaya, Indonesia.

Zitieren

APA

Santoso, A. R. B., Dradjat, R. S., Mustamsir, E., Sakti, S. P., Sananta, P., Sujuti, H., & Suroto, H. (2026). Comparison of immune responses to zirconia, polyether ether ketone (PEEK), and stainless-steel in orthopaedic implants. Medicinski Glasnik. https://doi.org/10.17392/2037-23-01

Vancouver

Santoso ARB, Dradjat RS, Mustamsir E, Sakti SP, Sananta P, Sujuti H, et al. Comparison of immune responses to zirconia, polyether ether ketone (PEEK), and stainless-steel in orthopaedic implants. Medicinski Glasnik. 2026. doi:10.17392/2037-23-01

BibTeX
@article{Santoso2026,
  author = {Santoso, Agung Riyanto Budi and Dradjat, Respati Suryanto and Mustamsir, Edi and Sakti, Setyawan Purnomo and Sananta, Panji and Sujuti, Hidayat and Suroto, Heri},
  title = {Comparison of immune responses to zirconia, polyether ether ketone (PEEK), and stainless-steel in orthopaedic implants},
  journal = {Medicinski Glasnik},
  year = {2026},
  doi = {10.17392/2037-23-01}
}
Kostenlos starten →